Recall Z-1191-2024

NxStage MDS Corporation · initiated January 29, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.

Codes / lots: UDI-DI M535CAR5350; Package DI M535CAR5355 Lots 20878001, 30978013, 30978032, 31178011, 31178019

Reason for recall

Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.

FDA-determined cause: Component design/selection

Action taken

The firm began notifying affected customers on January 29, 2024. Customers were informed that the issue may occur when the needleless access port located next to the filter has been accessed and there is a buildup of pressure in the circuit. Customers were instructed to inspect their inventory for affected product and quarantine recalled product. Customers should not use affected product and should ensure all users of the CAR-535 are aware of the notice. Customers should contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243) to arrange for the return of all affected products. This number may also be contacted if you have any questions or comments.

Details

Recall number
Z-1191-2024
Initiated
January 29, 2024
Posted by FDA
February 27, 2024
Product code
KDI – Dialyzer, High Permeability With Or Without Sealed Dialysate System
Quantity
693
Distribution
US domestic distribution to Dialysis Centers located in AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI. No international distribution.
Recalling firm
NxStage MDS Corporation, Lawrence, MA

Official FDA recall record