Recall Z-1195-2017
Argon Medical Devices, Inc · initiated January 18, 2017
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
T-LOK BONE MARROW BIOPSY NEEDLE 8G x 4 EXT CAN; T-LOK BONE MARROW BIOPSY NEEDLE 11G x 4 EXT CAN Intended for the purpose of harvesting bone and/or bone marrow specimens
Codes / lots: lots: 11157523, 11158179, 11158180, 11159101, 11159729
Reason for recall
The recalling firm received a complaint from one of their customers of a potential defect in the forming of the primary package of the blister that may compromise the sterility of the product inside the tray.
FDA-determined cause: Equipment maintenance
Action taken
Argon Medical Devices, Inc. is sending an Urgent Product Recall Notice letter dated January 17, 2017, by registered mail to all affected customers. Customers were asked to determine if their inventory shows the package defect and discontinue use of the device and promptly return the affected product. For questions regarding this recall call 903-403-0749.
Details
- Recall number
- Z-1195-2017
- Initiated
- January 18, 2017
- Posted by FDA
- February 13, 2017
- Terminated
- May 17, 2017
- Product code
- KNW – Instrument, Biopsy
- Quantity
- 772 boxes
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Switzerland, Canada, Australia, New Zealand, Ecuador, Colombia, and Japan.
- Recalling firm
- Argon Medical Devices, Inc, Athens, TX
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1194-2017 Class II | BONE MARROW ASPIRATION NEEDLE 15G x 2.688 MAX and 15G x 4 MAX Intended for the purpose…The recalling firm received a complaint from one of their customers of a potential defect in the forming of… | 01/18/2017Terminated |
| Z-1196-2017 Class II | HARVEST NEEDLE 15GA X 2 Intended for the purpose of harvesting bone and/or bone marrow…The recalling firm received a complaint from one of their customers of a potential defect in the forming of… | 01/18/2017Terminated |