Recall Z-1221-2024

Ortho-Clinical Diagnostics, Inc. · initiated January 22, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS 5600 Integrated System (New and Refurbished)

Codes / lots: UDI/DI: (New) 10758750002740, (Refurbished) 10758750007110; Serial Numbers: J56000280, J56000330, J56000415, J56000631, J56000634, J56000721, J56000839, J56000871, J56001088, J56001209, J56001228, J56001265, J56001305, J56001355, J56001834, J56001855, J56001941, J56001998, J56002111, J56002239, J56002476, J56002492, J56002531, J56002561, J56002649, J56002655, J56002658, J56002691 and J36001164 through J36001692 (inclusive).

Reason for recall

Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS 3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems which may lead to erroneous test results.

FDA-determined cause: Component change control

Action taken

An URGENT PRODUCT CORRECTION NOTIFICATION letter dated 01/22/24 was sent to customers. REQUIRED ACTIONS: To reduce the risk of unintended sample aspiration, refer to the procedure provided in this notification titled, Temporary TRAY Loading Instructions . QuidelOrtho recognizes that the Temporary TRAY Loading Instructions procedure may not be tenable for all laboratories due to testing volumes. Review your laboratory risk management plan(s) to assess the need for this further mitigation. Acknowledge your understanding of this notification by completing the enclosed Confirmation of Receipt form no later than January 30, 2024. Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System until this issue has been resolved. If your laboratory has experienced the anomaly described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization (formerly Ortho Care). QuidelOrtho has identified root cause however, our investigation is still ongoing. We are currently working on a Hardware Modification (MOD) to resolve this issue and will communicate again once the MOD or additional information is available. Current estimated availability date of the MOD is 2nd quarter 2024. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.

Details

Recall number
Z-1221-2024
Initiated
January 22, 2024
Posted by FDA
March 1, 2024
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
1460 units (506 US, 954 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.
Recalling firm
Ortho-Clinical Diagnostics, Inc., Rochester, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1220-2024 Class IIVITROS 3600 Immunodiagnostic System (New and Refurbished)Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS 3600… 01/22/2024Open, Classified
Z-1222-2024 Class IIVITROS XT7600 Integrated SystemPotential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS 3600… 01/22/2024Open, Classified