Recall Z-1227-2024

Ortho-Clinical Diagnostics, Inc. · initiated January 26, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Codes / lots: VITROS System Software Version 3.8.1 UDI-DI 10758750031986 Serial Numbers US: J34500180 J34500289 J34500422 J34500473 J34500622 J34500861 J34500887 OUS: J34500515 J34500609 J34500802 J34500940

Reason for recall

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

FDA-determined cause: Software Design Change

Action taken

On January 26, 2024, the firm sent a customer letter to all customers who have V3.8.1 installed on their VITROS Analyzer. Customers were informed of this issue and instructed to follow the instructions when using the VITROS System onboard QC package to update the QC baseline. The software will be updated to mitigate both issues within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.

Details

Recall number
Z-1227-2024
Initiated
January 26, 2024
Posted by FDA
March 1, 2024
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
7 US; 4 OUS
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.
Recalling firm
Ortho-Clinical Diagnostics, Inc., Rochester, NY

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1224-2024 Class IIVITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914…A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline… 01/26/2024Open, Classified
Z-1225-2024 Class IIVITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline… 01/26/2024Open, Classified
Z-1226-2024 Class IIVITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline… 01/26/2024Open, Classified
Z-1228-2024 Class IIVITROS XT 7600 Integrated System, Catalog No. 6844461A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline… 01/26/2024Open, Classified