Recall Z-1229-2007
C.R. Bard Access Systems, Inc · initiated August 10, 2007
Product
Guidewire, 70cm x 1mm, REF 5583705, Sterile, Single Use, BARD Access Systems, Hemodialysis Catheter, Salt Lake City, UT 84116.
Codes / lots: Product Code 5583705, Lot Numbers: 22FMA202, 22IMO531, 22KMO573, 22DN2914, 22FN5232, 22GN5835, 22GN5878, 22HN6574, 22BO6078, 22BO6163, 22FO6086, 22GO6026, 22HO6101, 22IO6111, 22KO6124, 22CP6268, 22EP6831, REPG0035, REPG0643, REPH0954, REPK0058, REPK0729, REPL0298, REQB0030, REQB0224, REQD0200, REQF0429, REQF0736, REQJ0282, REQK0875, RERC0052, RERD0719.
Reason for recall
Metal guidewires used in Hemodialysis Catheters may have been reverse loaded within the dispenser. Potential exists for the stiff end of the guidewire to be inserted into the patient instead of the flexible end.
FDA-determined cause: Other
Action taken
Consignees were notified by letter on 08/10/2007 and asked to remove and return all unused affected product to Bard.
Details
- Recall number
- Z-1229-2007
- Initiated
- August 10, 2007
- Posted by FDA
- September 20, 2007
- Terminated
- April 25, 2008
- Product code
- MPB – Catheter, Hemodialysis, Non-Implanted
- Quantity
- 7225 units
- Distribution
- Worldwide, including USA, Canada, Latin America, and Europe (Austria, Belgium, Finland, France, Germany, Great Britain, Holland, Italy, Norway, Spain, and Sweden).
- Recalling firm
- C.R. Bard Access Systems, Inc, Salt Lake City, UT