Recall Z-1234-2025

Scientia Vascular, Inc. · initiated January 2, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke

Codes / lots: Catalog Number: SC038-127-001 UDI-DI code: 00818075011025 Lot Numbers: 030471 030317 030345 030376 030427 030451

Reason for recall

Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.

FDA-determined cause: Under Investigation by firm

Action taken

On 01/02/2025, the firm initially notified customer via email an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that through internal evaluation of finished goods of the Socrates 38 Aspiration Catheter, one lot of pouches containing the 127 cm Socrates 38 Aspiration Catheter was identified with pouches that had a gap in the sterile barrier seal. This was a result of the pouch not being centered with the sealing bar during the sealing process. On 01/10/2025, The firm sent an updated letter informing customers of additional product lots were added to the Recall. Customers are instructed to: Segregate Recalled Product. Immediately remove all impacted (see list above) Socrates 38 Catheter(s) from their inventory (regardless of its location) and segregate this product in a secure location for return to Scientia Vascular, Inc. In situations where a device has already been used in a patient procedure, closely monitor patients for signs and symptoms of post-surgical infection. Customer Service will be available to provide additional information and provide replacement orders at customercare@scientiavascular.com or (888)385-9016; Monday through Friday, 8:00 AM to 4:30 PM, Mountain Time.

Details

Recall number
Z-1234-2025
Initiated
January 2, 2025
Posted by FDA
February 26, 2025
Product code
NRY – Catheter, Thrombus Retriever
Quantity
120 devices
Distribution
U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.
Recalling firm
Scientia Vascular, Inc., West Valley City, UT

Official FDA recall record