Recall Z-1242-2013

Smiths Medical ASD, Inc. · initiated March 28, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Wallace Sure-Pro Two Stage Embryo Replacement Catheter with Obturator PPB623 Sure-Pro Embryo Replacement Catheters are used for the introduction of embryos into the uterine cavity following in vitro fertilization. Sure-Pro Ultra Embryo Replacement Catheters have the added benefit of being visible under ultrasound. These products can be used in either a one- stage or two-stage procedure. In the one-stage procedure, the clinician pushes the inner catheter through the outer sheath and loads the embryo prior to performing the embryo transfer into the uterus. In the two-stage procedure, the clinician places the outer sheath through the cervical canal into the uterus, then loads the embryo in the inner catheter, and advances the inner catheter through the outer sheath to perform the embryo transfer into the uterus.

Codes / lots: Lot codes: 2000100 2025750 2044018 2044019 2065655 2120952 2152589 2152590 2156664 2156665 2170927 2295308 2305235 2305236 2305237 2305238 2308139 2308141 2313280 2313281 2320238 2324603 2336961 2341418 2359095 2363587 2363588 2363589 2363592 2363593 2363594 2363595 2363596 2363597 2363598

Reason for recall

Unable to pass catheter through outer sheath during embryo transfer

FDA-determined cause: Process control

Action taken

The firm, Smiths Medical, sent an "URGENT FIELD SAFETY NOTICE" dated March 28, 2013 to its customers. The notification describes the product, problem and actions to be taken. The customers were provided with two options for proceeding: 1) continue to use product if the product passes a pre-test; or 2) return affected product to Smiths Medical for replacement or credit. In addition, the customers were instructed to inspect their inventory and quarantine all affected products; perform a count of affected products currently in inventory; complete and return the attached Confirmation Form via fax to +1-866-258-6068 or email to pamela.harrell@smiths-medical.com; send a copy of this notice to your customers identified having received affected product; and provide email confirmation of your customers notifications to Smiths medical at pamela.harrell@smiths-medical.com. Customers should report any issues with these products to Smiths Medicals Global Complaint Department at +1-866-216-8806 or globalcomplaints@smiths-medical.com. If you should have any questions regarding this information, please contact Smiths Medical's Customer Service Department at +1-800-258-5361.

Details

Recall number
Z-1242-2013
Initiated
March 28, 2013
Posted by FDA
May 3, 2013
Terminated
May 19, 2021
Product code
MQF – Catheter, Assisted Reproduction
Quantity
37620 units
Distribution
Worldwide distribution: USA (Nationwide) and countries of: Arab Emirates (AE) Ireland (IE), Albania, Israel (IL), Argentina (AR),Italy (IT) Armenia (AM) , Japan (JP), Austria (AT) Jordan (JO), Bahrain (BH) Kuwait (KW) Belgium (BE), Lebanon (LB), Belarus , Netherlands (NL), Brazil (BR),Nigeria (NG), Bulgaria (BG) , Pakistan (PK), Canada (CA) Poland (PL), China (CN) Romania (RO), Columbia (CO)…
Recalling firm
Smiths Medical ASD, Inc., Keene, NH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1239-2013 Class IIWallace Sure-Pro/ Sure-Pro Ultra Embryo Replacement Catheters (with ultrasound) Product…Unable to pass catheter through outer sheath during embryo transfer 03/28/2013Terminated
Z-1240-2013 Class IIWallace Sure-Pro Ultra Embryo Replacement Catheter with Obturator Product Code: PEB623…Unable to pass catheter through outer sheath during embryo transfer 03/28/2013Terminated
Z-1241-2013 Class IIWallace Sure-Pro Ultra Embryo Replacement Catheter with Stylet Product Code:PES623…Unable to pass catheter through outer sheath during embryo transfer 03/28/2013Terminated
Z-1243-2013 Class IIWallace Sure-Pro Two Stage Embryo Replacement Catheter for difficult transfer with…Unable to pass catheter through outer sheath during embryo transfer 03/28/2013Terminated
Z-1244-2013 Class IIWallace Sure-Pro Single Stage Embryo Replacement Catheter Product Code: PP623 Sure-Pro…Unable to pass catheter through outer sheath during embryo transfer 03/28/2013Terminated