Recall Z-1257-2023
Beckman Coulter, Inc. · initiated January 26, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, DxC 600i Packaged(A25640).
Codes / lots: UDI number for the Access 2 instruments: 15099590265380. No UDI for Access 2 Section, DxC 60i Packaged, Catalog #A25640. PN 81600N serial numbers: 570512 570513 570514 570515 570516 570517 570518 570519 570520 570521 570522 570523 570524 570525 570526 570527 570528 570529 570530 570531 570532 570533 570534 570535 570536 570537 570538 570539 570540 570541 570542 570543 570544 570545 570546 570547 570548 570549 570550 570551 570552 570553 570554 570557 570559 570560 570562 570563 570564 570565 570566 570567 570568 570569 570570 570571 570572 570573 570574 570575 570576 570577 570578 570579 570580 570581 570582 570583 570584 570585 570586 570587 570588 570590 570591 570592 570593 570594 570598 570599 570600 570601 570602 570603 570604 570605 570606 570607 570608 570609 570610 570611 570612 570613 570614 570615 570616 570617 570618 570619 570620 570621 570622 570623 570624 570625 570626 570627 570628 570629 570630 570631 570632 570633 570634 570635 570636 570637 570638 570639 570640 570641 570642 570643 570644 570645 570646 570647 570648 570649 570650 570651 570652 570653 570654 570655 570656 570657 570658 570659 570660 570662 570664 570665 570666 570668 570669 570670 570671 570672…
Reason for recall
An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period.
FDA-determined cause: Component change control
Action taken
The recalling firm issued letters dated 1/19/2023 beginning 1/26/2023 via both email and first class mail. The letter explained the issue, which included photographs, impact, and the actions to be taken. The consignee was to inspect the wash arm lead screw nut band for damage following the instructions listed in Appendix B of the letter. This inspection was to be repeated during weekly maintenance until the affected nut band has been replaced. If the nut band appears to be intact and undamaged. the consignee may continue normal operation of the instrument. If Wash Arm Motion Errors are observed or if the nut band appears to be damaged in any way, specific instructions are provided. The recalling firm will replace the affected nut bands for all Access 2 Immunoassay Systems listed in Appendix A of the letter. A service representative will contact the consignee to schedule the wash arm lead screw nut band replacement. The information in the notification is to be shared with the laboratory staff and retained as part of the laboratory Quality System documentation. If the affected device has been forwarded to another laboratory, the consignee is requested to provide them a copy of the letter. Response to the notification is requested either electronically or manually using the enclosed response form within 10 days.
Details
- Recall number
- Z-1257-2023
- Initiated
- January 26, 2023
- Posted by FDA
- March 15, 2023
- Product code
- JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Quantity
- 4,137 devices (990 U.S. and 3,147 OUS)
- Distribution
- Worldwide and Nationwide Distribution. There was government/military distribution. Foreign distribution was made to Algeria, Andorra, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Bonaire, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, France…
- Recalling firm
- Beckman Coulter, Inc., Chaska, MN