Recall Z-1261-2025

Medtronic MiniMed, Inc. · initiated January 31, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)

Codes / lots: All Lot Numbers; All UDI numbers

Reason for recall

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

FDA-determined cause: Device Design

Action taken

On January 31, 2025, Medtronic MiniMed issued a "Urgent Medical Device Correction" notification to all affected consignees. The notification was distributed via: e-mail, certified mail, SMS text message or phone call. 1. Monitor Your Glucose Levels: 2. Discuss how to prepare for situations like this with your healthcare professional. 3. Respond to Alerts and Symptoms. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter by either completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://info.medtronicdiabetes.com/pump-delivery-volume] to acknowledge this communication.

Details

Recall number
Z-1261-2025
Initiated
January 31, 2025
Posted by FDA
March 3, 2025
Product code
OZP – Automated Insulin Dosing Device System, Single Hormonal Control
Quantity
100,074 units
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY and the countries of Algeria, Andorra, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh…
Recalling firm
Medtronic MiniMed, Inc., Northridge, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1255-2025 Class IIParadigm REF: MMT-712 and MMT-715All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1256-2025 Class IIParadigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1257-2025 Class IIParadigm insulin pump, REF: MMT-754All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1258-2025 Class IIMiniMed insulin pump, REF: 620G (MMT-1750)All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1259-2025 Class IIMiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1260-2025 Class IIMiniMed insulin pump, REF: 640G (MMT-1711, MMT-1712, MMT-1751, MMT-1752)All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1262-2025 Class IIMiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851)All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1263-2025 Class IIMiniMed insulin pump, REF: 720G (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1264-2025 Class IIMiniMed insulin pump, REF: 740G (MMT-1811, MMT-1812, MMT-1861, MMT-1862)All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1265-2025 Class IIMiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891…All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified
Z-1266-2025 Class IIMiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895…All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have… 01/31/2025Open, Classified