Recall Z-1264-2022

Olympus Corporation of the Americas · initiated April 20, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R

Codes / lots: UDI-DI: 04953170340833 Serial Number: 2836209F 2835978 2837272 2836299

Reason for recall

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

FDA-determined cause: Nonconforming Material/Component

Action taken

Olympus notified Urgent Medical Device Correction letter on 4/20/22 via email. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. The model and serial number can be found on the device as illustrated. 2. An Olympus representative will contact you to make arrangements for return of your affected device. Olympus will repair your unit free of charge. In the meantime, follow the instruction manual for inspection of the device before use. Should any irregularities be identified, including a loose parts, do not use the device and contact Olympus for repair. 3. Olympus will arrange for a service loaner in an effort not to disrupt clinical care. Contact me directly at 647-999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.

Details

Recall number
Z-1264-2022
Initiated
April 20, 2022
Product code
FGB – Ureteroscope And Accessories, Flexible/Rigid
Quantity
4 units
Distribution
CA FL MA MD MI NJ OH PR WI
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1262-2022 Class IIOlympus CYSTONEPHROFIBERSCOPE Model: CYF-5Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may… 04/20/2022Open, Classified
Z-1263-2022 Class IIOlympus URETERO-RENO FIBERSCOPE Model: URF-P6Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may… 04/20/2022Open, Classified