Recall Z-1286-2016

Ossur H / F · initiated February 24, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Knee Undersleeve Gray Product Usage: Knee undersleeves are worn as an accessory for comfort under knee braces.

Codes / lots: Item No: B-250150201 B-250150202 B-250150203 B-250150204 B-250150205 B-250150206

Reason for recall

Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar, Rebound Cartilage, Rebound Cartilage Custom, Rebound Air Walker, Rebound Diabetic Walker, Back Support Suspenders Clip on Black, Back Support Suspenders Sew On Black, Extreme Custom, Unloader Custom XT, Unloader Custom Lite, Unloader Custom, and Papoose Infant Spine Immobilizers because the p

FDA-determined cause: Labeling False and Misleading

Action taken

The firm sent out customer notification letters on 02/26/16. The notification letter instructed customers to include the latex caution notice in any existing inventory of affected products. The latex caution notice can be retrieved at the conclusion of the response survey or by going directly to safetyalerts.ossur.com. Should customers require ssur to provide pre-printed notices for existing inventory, contact ssur customer service. For the ssur Miami J Cervical Collars with lot numbers before MX150727 customers are instructed to remove and discard the assist strap, which contains natural rubber latex, located in the patient pack. Recipients of the notice with should take the following actions: 1. Pass the notice to those who need to be aware within the organization or to any organization where the affected devices have been transferred. 2. Maintain awareness on this notification for an appropriate period. 3. If have further distributed the products, identify customers and notify them at once of the product alert. It is recommended to include a copy of the alert notification letter. 4. If any of your customers are currently wearing a product, check for any skin irritations or other reactions consistent with natural rubber latex allergies. 5. To verify have received the notification, visit safetyalerts.ossur.com to complete the Response Survey. If any questions, contact ssur Customer Service at (800) 233-6263. The notification letter for the Papoose Infant Spine Immobilizer contained the following additional actions/instructions: Examine inventory, quarantine products subject to the recall and contact customer service for a return authorization to return the product to ssur. If any of customers are currently wearing a product, it is recommended replacing the unit with one shipped after the transition lot number MX150727. Check for any skin irritations or other reactions consistent with natural rubber latex allergies.

Details

Recall number
Z-1286-2016
Initiated
February 24, 2016
Terminated
August 15, 2016
Product code
ITQ – Joint, Knee, External Brace
Quantity
456
Distribution
U.S. distribution to the following; AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, RI, TN, TX, UT, VA, VT, WA, WI, WV
Recalling firm
Ossur H / F, Reykjavik

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1285-2016 Class IIKnee Shield (Patella Cup) Product Usage: Knee Shield patella cup is an accessory used…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1287-2016 Class IINecLoc Extrication Collar Product Usage: Necloc collars are applied to patients…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1288-2016 Class IIMiami J Collar Product Usage: Miami J collars are applied to patients afflicted with…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1289-2016 Class IIRebound Cartilage Product Usage: Rebound Cartilage is an external knee brace that…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1290-2016 Class IIRebound Cartilage Custom Product Usage: Rebound Cartilage Custom is an external knee…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1291-2016 Class IIRebound Air Walker Product Usage: Rebound Air Walkers are indicated for grade 2 and…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1292-2016 Class IIRebound Diabetic Walker Product Usage: Rebound Diabetic Walkers are used for the…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1293-2016 Class IIBack Support Suspenders Clip on Black Back Support Suspenders Sew on BlackOssur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1294-2016 Class IIUnloader XT Product Usage: Unloader XT Extreme is an external knee braces to provide…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1295-2016 Class IIUnloader Custom Lite Product Usage: Unloader Custom Lite is an external knee braces to…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1296-2016 Class IIUnloader Custom Product Usage: Unloader Custom is an external knee braces to provide…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1297-2016 Class IIExtreme Custom Product Usage: Extreme Custom is an external knee braces to provide…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated
Z-1298-2016 Class IIPapoose Infant Spine Immobilizer Item # Item Description: PA-100 Papoose PAR-100 Papoose…Ossur Americas is recalling , Knee Shield / Patella Cup, Knee Undersleeve Gray, NecLoc Extrication Collar… 02/24/2016Terminated