Recall Z-1292-2015
Baxter Healthcare Corp. · initiated October 16, 2013
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
The HomeChoice Pro APD System is a peritoneal dialysis system. The HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Codes / lots: Product Code: 5C8310; Serial Numbers: 80415, 680416, 680418, 680420, 680421, 680422, 680424, 680426, 680427, 680428, 680429, 680430, 680438, 680439, 680440, 680441, 680444, 680445, 680448, 680449, 680452, 680453, 680454, 680455, 680456, 680457, 680459, 680460, 680461, 680462, 680463, 680465, 680467, 680467, 680469, 680470, 680472, 680473, 680475, 680476, 680477, 680478, 680480, 680481, 680482, 680463, 680484, 680485, 680486, 680489, 680490, 680491, 680494, 680495, 680496, 680498, 680499, 680500, 680501, 680502, 680503, 680504, 680505, 680506, 680507, 680508, 680509, 680510, 680516, 680517, 680519, 680520, 680521, 680522, 680524, 680525, 680526, 680527, 680527, 680528, 680529, 680530, 680532, 680533, 680534, 680536, 680537, 680538, 680539, 680541, 680542, 680543, 680547, 680552, 680553, 680544, 680555, 680561, 680562, 680566, 680567, 680585, 680586, 680587, 680588, 680589, 680590, 680591, 680593, 680594, 680595, 680596, 680597, 680598, 680599, 680600, 680601, 680602, 680603, 680604, 680605, 680606, 680607, 680608, 680609, 680611, 680612, 680613, 680614, 680615, 680616, 680617, 680618, 680619, 680620, 680621, 680622, 680623, 680624, 680625, 680626, 680627, 680628, 680629, 680630…
Reason for recall
The occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the back plate and cause Slow / No Flow alarm requiring therapy termination.
FDA-determined cause: Process design
Action taken
An Urgent Device Correction letter dated 10/16/13 was mailed to dialysis centers and medical facilities. A similar Urgent Device Correction letter dated 10/31/13 was mailed to dialysis home patients. The letters requested for customers to examine their devices to determine if the devices are being recalled. If the devices are on the attached list of recalled devices, the customer is to contact Baxter Technical Services at 1-800-553-6898 (7 days a week, 24 hours a day) to arrange for a swap cycler. The letter further requests that customers complete the attached Customer Reply Form and return it to Baxter by fax or e-mail.
Details
- Recall number
- Z-1292-2015
- Initiated
- October 16, 2013
- Posted by FDA
- March 18, 2015
- Terminated
- July 21, 2017
- Product code
- FKX – System, Peritoneal, Automatic Delivery
- Quantity
- 75 units (US)
- Distribution
- Worldwide Distribution-US (nationwide) including the states of AK, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, VT, WA, and WI, and the countries of Australia, Canada, and Mexico.
- Recalling firm
- Baxter Healthcare Corp., Round Lake, IL