Recall Z-1297-2021

Stryker Communications · initiated January 29, 2021

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good Model Number o P38198 - ASM, MFR CONTROLLER, FIXED STANDARD o P38199 - ASM, MFR CONTROLLER, LIFT STANDARD o P40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD o P40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD o P46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD

Codes / lots: Serial Numbers: 2258046 2863490

Reason for recall

The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.

FDA-determined cause: Under Investigation by firm

Action taken

An Urgent Customer Notification Letter was mailed (certified) to the affected consignees on 01/29/2021. Customers will be notified by letter and a Stryker Representative will visit each account for product inspection and replacement. If customers notice any issues prior to inspection, customers should discontinue use and call Technical Support immediately at (866) 841-5663.

Details

Recall number
Z-1297-2021
Initiated
January 29, 2021
Product code
FQO – Table, Operating-Room, Ac-Powered
Quantity
1,393 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.
Recalling firm
Stryker Communications, Flower Mound, TX