Recall Z-1299-2013

Integra LifeSciences Corporation · initiated March 28, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BioMend Extend" Absorbable Collagen Membrane BioMend Extend" Absorbable Collagen Membrane is an absorbable, implantable material that is indicated for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus.

Codes / lots: 1104846, 1105175 1105253, 1110810 1110811, 1111149 1111859, 1104847 1105176, 1105254 1110809, 1111860 1104848, 1110819 1110820, 1111148 1111861

Reason for recall

Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.

FDA-determined cause: Process control

Action taken

The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: FCA@integralife.com or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product. "However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol." For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723.

Details

Recall number
Z-1299-2013
Initiated
March 28, 2013
Posted by FDA
May 10, 2013
Terminated
June 3, 2015
Product code
LYC – Bone Grafting Material, Synthetic
Quantity
20,032 boxes
Distribution
Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.
Recalling firm
Integra LifeSciences Corporation, Plainsboro, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1282-2013 Class IIDural Graft Matrix DuraGen Dural Graft Matrix is an absorbable implant for repair of…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1283-2013 Class IISuturable DuraGen Dural Regenerative Matrix Suturable DuraGen Dural Regenerative Matrix…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1284-2013 Class IIDuraGen Plus Dural Regeneration Matrix DuraGen Plus Adhesion Barrier Matrix is an…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1285-2013 Class IIDuraGen XS Dural Regeneration Matrix DuraGen XS Dural Regenerative Matrix is an…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1286-2013 Class IINeuraWrap Nerve Protector NeuraWrap Nerve Protector is indicated for the management of…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1287-2013 Class IINeuraGen Nerve Guide NeuraGen Nerve Guide is indicated for repair of peripheral nerve…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1288-2013 Class IIINSTAT - Agent, Absorbable Hemostatic, Collagen Based INSTAT MCH is recommended for use…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1289-2013 Class IIHelistat Absorbable Hemostatic,Collagen Based Helistat is indicated in surgical…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1290-2013 Class IIHelitene Absorbable Collagen Hemostatic Sponge Helitene is indicated in surgical…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1291-2013 Class IIColla Tape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1292-2013 Class IICollaCote Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1293-2013 Class IICollaPlug Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1294-2013 Class IIHeliTape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1295-2013 Class IIHeliPlug Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1296-2013 Class IIOraTape Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1297-2013 Class IIOra Plug Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1298-2013 Class IIOraMem Absorbable Collagen Membrane OraMem Absorbable Collagen Membrane is an…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1300-2013 Class IILayershield Adhesion Barrier Matrix Layershield Adhesion Barrier Matrix is an absorbable…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1301-2013 Class IIBioMend Absorbable Collagen Membrane BioMend Absorbable Collagen Membrane is an…Due to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated
Z-1302-2013 Class IIAbsorbable Collagen Sponge Provided As Intermediate to Another CompanyDue to the process deviation, product lots in question may have been released with higher levels of pyrogens… 03/28/2013Terminated