Recall Z-1304-2025

CareFusion 303, Inc. · initiated February 18, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software

Codes / lots: Software v12.5.1 or v12.5.2, and prior versions (4.33, 12.0.1, 12.0.2, 12.1, 12.1.2, 12.3) UDIs: 10885403960123 10885403519666 (10885403960116)

Reason for recall

Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).

FDA-determined cause: Under Investigation by firm

Action taken

On February 18, 2025, Becton, Dickinson issued a " Urgent Medical Device Product Correction" Notification. to affected consignees via Fed-Ex and E-Mail. 1. Users should review all infusion parameters pre-populated and confirm they are correct before pressing START as stated in the BD Alaris User Manual. 2. Disseminate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 3. BD Project Managers will engage with you to address any questions regarding this notice, and specific next steps for your facility. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form, regardless of whether you have any of the impacted material. This will allow BD to confirm receipt of this notification. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.

Details

Recall number
Z-1304-2025
Initiated
February 18, 2025
Posted by FDA
March 13, 2025
Product code
PHC – Infusion Safety Management Software
Distribution
US: AZ, CA, DC, FL, GA, IL, IN, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI OUS: None
Recalling firm
CareFusion 303, Inc., San Diego, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1305-2025 Class IBD Care Coordination Engine (CCE) Infusion Adapter, Safety Management SoftwareSoftware issue that may result in outdated automated programming request(APR) being sent to the progressive… 02/18/2025Open, Classified