Recall Z-1306-2026

Fujirebio Diagnostics, Inc. · initiated December 11, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Product Name: Lumipulse G -Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G -Amyloid 1-42-N Plasma Calibrators: The -Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G -Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 1.5 mL (4 Concentrations) CAL 1 0 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL -Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL -Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Codes / lots: Model/Catalog Number: 235454; UDI-DI: 04987270235454; Lot Numbers: D8C6022U, D8C6044U;

Reason for recall

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

FDA-determined cause: Under Investigation by firm

Action taken

On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.

Details

Recall number
Z-1306-2026
Initiated
December 11, 2025
Posted by FDA
February 5, 2026
Product code
SET – Immunoassay Blood Test For Amyloid Pathology Assessment
Quantity
128 units
Distribution
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Recalling firm
Fujirebio Diagnostics, Inc., Malvern, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1301-2026 Class IIProduct Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number:…Customers may experience inaccurate test results by observing a higher frequency of classification in the… 12/11/2025Open, Classified
Z-1302-2026 Class IIProduct Name: Lumipulse G pTau 217 Plasma Immunoreaction Cartridges Model/Catalog…Customers may experience inaccurate test results by observing a higher frequency of classification in the… 12/11/2025Open, Classified
Z-1303-2026 Class IIProduct Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557…Customers may experience inaccurate test results by observing a higher frequency of classification in the… 12/11/2025Open, Classified
Z-1304-2026 Class IIProduct Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software…Customers may experience inaccurate test results by observing a higher frequency of classification in the… 12/11/2025Open, Classified
Z-1305-2026 Class IIProduct Name: Lumipulse G -Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog…Customers may experience inaccurate test results by observing a higher frequency of classification in the… 12/11/2025Open, Classified
Z-1307-2026 Class IIProduct Name: Lumipulse -Amyloid Plasma Controls Model/Catalog Number: 81559 Software…Customers may experience inaccurate test results by observing a higher frequency of classification in the… 12/11/2025Open, Classified