Recall Z-1315-2024
Mckesson Medical-Surgical Inc. Corporate Office · initiated January 9, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Abbott Point of Care Control Control i-STAT Level 3 1.7 mL
Codes / lots: Catalogue Number: 06F1401; UDI/DI: 54749000661; Lot No: 121164;
Reason for recall
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
FDA-determined cause: Storage
Action taken
McKesson notified customers by phone January 9, 2024 with a follow-up call January 17, 2024 instructing customers of three scenarios based upon the products they received and the investigation in conjunction with relevant manufacturer. The firm also sent Urgent Product Recall Notifications via email on February 8, 2024 instructing customers to immediately examine inventory and quarantine all affected devices, complete and return the provided response form to MMSQRCProductReviews@Mckesson.com, and notify customers who you have further distributed recalled devices to of this recall and to segregate and quarantine any devices they have on hand. Replacement products will be provided and may have the same lot number(s) as the recalled items but are not affected by the recall and are acceptable for use. Customers with questions can contact McKesson directly by phone at (800) 688-8840. Representatives are available Monday to Friday, 8:00 am to 5:00 pm PST.
Details
- Recall number
- Z-1315-2024
- Initiated
- January 9, 2024
- Posted by FDA
- March 19, 2024
- Product code
- JJY – Multi-Analyte Controls, All Kinds (Assayed)
- Quantity
- 2 units
- Distribution
- US Nationwide distribution in the states of AZ, PA and TX.
- Recalling firm
- Mckesson Medical-Surgical Inc. Corporate Office, Richmond, VA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1311-2024 Class II | Sekisui Diagnostics Control Kit OSOM BVBLUE Bacterial VaginosisFacility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures… | 01/09/2024Open, Classified |
| Z-1312-2024 Class II | Abbott Rapid Dx N America LLC Reproductive Endocrinology Control Set McKesson Consult"…Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures… | 01/09/2024Open, Classified |
| Z-1313-2024 Class II | Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures… | 01/09/2024Open, Classified |
| Z-1314-2024 Class II | Abbott Point of Care Control Control i-STAT Level 1 10 X 1.7 mLFacility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures… | 01/09/2024Open, Classified |