Recall Z-1317-2008

Lifescan Inc · initiated December 18, 2007

Terminated

Product

LifeScan OneTouch Data Management Software Version 1.0, Part Number: 021-722-01(for OneTouch Select and UltraMini (Download) Blood Glucose Meters), LifeScan Inc, Milpitas, CA 95037.

Codes / lots: Part Number: 021-722-01

Reason for recall

Meter Temporarily Freezes-- A software compatibility issue may cause the blood glucose meter to cease operations and freeze temporarily.

FDA-determined cause: Software design

Action taken

The firm sent an Urgent Recall letter to Health Care Professionals on 12/18/2007. The letter advised user to uninstall software drivers and provided options on how to uninstall. For further assistance, please contact LifeScan Customer Service at 1-866-679-9250.

Details

Recall number
Z-1317-2008
Initiated
December 18, 2007
Posted by FDA
April 8, 2008
Terminated
July 18, 2008
Product code
NBW – System, Test, Blood Glucose, Over The Counter
Quantity
260 units
Distribution
Nationwide Distribution including the states of AL, CA, IN, IO, MA, MD, MI, MS, NE, NY, NC, OR, PA, TN, TX, UT, VT, VA, WV, and WI.
Recalling firm
Lifescan Inc, Milpitas, CA

Official FDA recall record