Recall Z-1317-2018

Smith & Nephew, Inc. · initiated March 7, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

Codes / lots: Lot Number 17GM04522

Reason for recall

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

FDA-determined cause: Under Investigation by firm

Action taken

The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.

Details

Recall number
Z-1317-2018
Initiated
March 7, 2018
Posted by FDA
March 20, 2018
Terminated
October 31, 2018
Product code
HWC – Screw, Fixation, Bone
Quantity
59 units
Distribution
US, Puerto Rico, Malaysia
Recalling firm
Smith & Nephew, Inc., Memphis, TN