Recall Z-1317-2020

Nobel Biocare Usa Llc · initiated March 14, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Implant Retrieval Instrument Hex & Tri-Chi NP/RP 31mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.

Codes / lots: (01)07332747083774(10)73303 Catalog# 37472 & Tri-Chi

Reason for recall

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

FDA-determined cause: Packaging process control

Action taken

On 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees: 1. Identify affected devices as detailed in the Customer Reply Form. 2. Do not use affected devices. 3. Send back the Customer Reply Form and the affected devices. The firm will replace affected devices free of charge. This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. We kindly ask you to follow instructions provided in this Recall Letter and return the attached Customer Reply Form to the firm via email or fax at 714- 282-5073.

Details

Recall number
Z-1317-2020
Initiated
March 14, 2019
Posted by FDA
February 19, 2020
Terminated
February 15, 2024
Product code
NDP – Accessories, Implant, Dental, Endosseous
Quantity
197 units
Distribution
Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
Recalling firm
Nobel Biocare Usa Llc, Yorba Linda, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1302-2020 Class IIGuided Start Drills/Counterbore NobRpl NP - Product Usage: The Guided Drills (Twist…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1303-2020 Class IIDrill Tapered NP 3.5x8 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1304-2020 Class IIDense Bone Drill 6.0 6x16 mm Single-Pat- Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1305-2020 Class IIScrew Tap Tapered WP Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1306-2020 Class IIDrill Tapered NP 3.5x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1307-2020 Class IIDrill Tapered NP 3.5x11.5 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1308-2020 Class IIDrill Tapered RP 4.3x10 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1309-2020 Class IIDrill Tapered RP 4.3x11.5 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1310-2020 Class IIDrill Tapered RP 4.3x13 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step)…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1311-2020 Class IIDrill Tapered WP 5x10 mm Single-Pat - Product Usage: Drills are intended to be used in…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1312-2020 Class IIDrill Tapered WP 5x11.5 mm Single-Pat - Product Usage: Drills are intended to be used in…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1313-2020 Class IIDrill Tapered 6.0 6x8 mm Single-Pat - Product Usage: Drills are intended to be used in…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1314-2020 Class IIDense Bone Drill NP 3.5x16 mm Single-Pat - Product Usage: Drills are intended to be used…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1315-2020 Class IIDense Bone Drill WP 5x13 mm Single-Pat - Product Usage: Drills are intended to be used…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1316-2020 Class IIImplant Retrieval Instrument Hex & Tri-Ch NP/RP 22mm - Product Usage: They can be used…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1318-2020 Class IIImplant Retrieval Instrument CC 3.0 31 mm- Product Usage: They can be used to remove…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1319-2020 Class IIImplant Retrieval Instrument CC NP & Ext Hex WP 22 mm - Product Usage: They can be used…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1320-2020 Class IIImplant Retrieval Instrument CC RP & Tri Ch Hex WP 22 mm - Product Usage: They can be…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1321-2020 Class IIImplant Retrieval Instrument Tri Channel 6.0 22 mm - Product Usage: They can be used to…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1322-2020 Class IIAbutment Screw Retrieval Instrument 3.0/NP - Product Usage: The instruments for abutment…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1323-2020 Class IIImplant Retrieval Kit - Product Usage: The kit itself does not have a specific intended…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1324-2020 Class IIAbutment Screw Retrieval Kit - Product Usage: The kit itself does not have a specific…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated
Z-1325-2020 Class IIAbutment Screw Retrieval Instrument RP/WP/6.0 - Product Usage: The instruments for…Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be… 03/14/2019Terminated