Recall Z-1317-2022

Covidien Llc · initiated June 8, 2022

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Palindrome H Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.

Codes / lots: Code Information organized as "Model No., Serial No., GTIN": 8888119365, 1925200097, 20884521012810; 8888119371, 2031800117, 20884521056562; 8888123405, 2028200168, 20884521012940; 8888123405, 2127200428, 20884521012940; 8888145043, 1932200082, 10884521013223; 8888145043, 2003700042, 10884521013223; 8888145043, 2014900147, 10884521013223; 8888145043, 2023300169, 10884521013223; 8888145043, 2031800113, 10884521013223; 8888145043, 2034400185, 10884521013223; 8888145043, 2104600104, 10884521013223; 8888145043, 2108100054, 10884521013223; 8888145043, 2127200539, 10884521013223; 8888145043, 1919700110, 20884521013220; 8888145043, 1929500076, 20884521013220; 8888145043, 1932200082, 20884521013220; 8888145043, 2000700103, 20884521013220; 8888145043, 2003700042, 20884521013220; 8888145043, 2014900147, 20884521013220; 8888145043, 2019200001, 20884521013220; 8888145043, 2023300169, 20884521013220; 8888145043, 2028200174, 20884521013220; 8888145043, 2031800112, 20884521013220; 8888145043, 2031800113, 20884521013220; 8888145043, 2034400185, 20884521013220; 8888145043, 2034400187, 20884521013220; 8888145043, 2104600100, 20884521013220; 8888145043, 2104600104, 20884521013220; 8888145043…

Reason for recall

There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

FDA-determined cause: Process control

Action taken

Medtronic distributed a recall notification to customers by mail on 06/08/2022. Customers were instructed to immediately quarantine and discontinue use for specific lots of chronic hemodialysis catheters, which are to be returned to Medtronic. If customers have further distributed product, they are asked to forward the information from the recall notification to those recipients. All customers are asked to complete and return the Consignee Product Retrieval Confirmation Form to rs.gmbfcamitg@medtronic.com. Additionally, consignees are asked to share the notice within their organization and retain it for their records regarding the retrieval of unused product. If a consignee is aware of any adverse events associated with recalled product, they are to submit a report to FDA via Medwatch. Customers with any questions can contact their Medtronic Representative or call Customer Service at 800-962-9888, option 2. ***Updated 8/16/22*** Medtronic has created a website to help customers identify affected product: https://www.medtronic.com/covidien/en-us/products/c/renal-care-recall-lookup-tool.html#lotLookupTool

Details

Recall number
Z-1317-2022
Initiated
June 8, 2022
Posted by FDA
July 7, 2022
Product code
MSD – Catheter, Hemodialysis, Implanted
Quantity
26,474 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Brunei…
Recalling firm
Covidien Llc, Mansfield, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1312-2022 Class IPalindrome Precision Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1313-2022 Class IPalindrome Precision H Chronic Catheter, Sterile, Single Use, Rx Only. For vascular…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1314-2022 Class IPalindrome Precision SI Chronic Catheter, Sterile, Single Use, Rx Only. For vascular…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1315-2022 Class IPalindrome Precision HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1316-2022 Class IPalindrome Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1318-2022 Class IPalindrome SI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1319-2022 Class IPalindrome HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified
Z-1320-2022 Class IMahurkar Chronic Carbothane (Maxid) Catheter, Sterile, Single Use, Rx Only. For acute…There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an… 06/08/2022Open, Classified