Recall Z-1328-2016

MicroAire Surgical Instruments, LLC · initiated February 8, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MicroAire PAL Cannula Product Usage - Power Aspiration Device

Codes / lots: Include lot numbers/serial numbers 0913228216, 0414243005, 0215266350, 0415269875, 0515271334

Reason for recall

MicroAire has initiated a recall on product PAL-R4011XL (40cm x 4mm Del Vecchio French Rasp cannula), due to our receipt of three (3) complaints that it bent or broke during surgery. Accordingly, the PAL-R4011XL has also been put on production hold until further notice.

FDA-determined cause: Nonconforming Material/Component

Action taken

The firm is sending letters to all customers who have received PAL-R4011XL from August 2013- October 2015. The firm sent these letters by UPS to expedite the delivery and for traceability.

Details

Recall number
Z-1328-2016
Initiated
February 8, 2016
Terminated
February 14, 2017
Product code
GEA – Cannula, Surgical, General & Plastic Surgery
Quantity
145 manufactured and distributed
Distribution
Worldwide Distribution - US Nationwide
Recalling firm
MicroAire Surgical Instruments, LLC, Charlottesville, VA