Recall Z-1335-2019

Shippert Medical Technologies · initiated February 2, 2018

Class II Completed

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Tissu-Trans MEGA 1500, Catalog 3-TT-MEGA 1500, sterile, disposable, one-patient use. The firm name on the label is Shippert Medical Technologies, Centennial, CO.

Codes / lots: 61471, 61625, 61754, 61840, 61945, 61950, 62135, 62137, and 62157

Reason for recall

The pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.

FDA-determined cause: Packaging process control

Action taken

Letters dated 1/9/2018 were issued beginning 2/2/2018 via registered mail and Fed Ex explaining the reason for recall, risk to health, description of how to recognize the device failed, and actions to be taken.

Details

Recall number
Z-1335-2019
Initiated
February 2, 2018
Product code
MUU – System, Suction, Lipoplasty
Quantity
6,553 units were distributed for all catalog numbers
Distribution
Distribution was nationwide, including PR. There was government/military distribution. Foreign distribution was made to Australia, Austria, Belgium, Denmark, Germany, Greece, India, Italy, Japan, Netherlands, New Zealand, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, and United Kingdom,
Recalling firm
Shippert Medical Technologies, Centennial, CO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1330-2019 Class IITissu-Trans FILTRON 100, Catalog 3-TT-FILTRON 100, sterile, disposable, one-patient use.…The pouch integrity of the sterile product can't be assured without a visual examination due to possibly… 02/02/2018Completed
Z-1331-2019 Class IITissu-Trans FILTRON 250, Catalog 3-TT-FILTRON 250, sterile, disposable, one-patient use.…The pouch integrity of the sterile product can't be assured without a visual examination due to possibly… 02/02/2018Completed
Z-1332-2019 Class IITissu-Trans FILTRON 500, Catalog 3-TT-FILTRON 500, sterile, disposable, one-patient use.…The pouch integrity of the sterile product can't be assured without a visual examination due to possibly… 02/02/2018Completed
Z-1333-2019 Class IITissu-Trans FILTRON 1000, Catalog 3-TT-FILTRON 1000, sterile, disposable, one-patient… 02/02/2018Completed
Z-1334-2019 Class IITissu-Trans FILTRON 2000, Catalog 3-TT-FILTRON 2000, sterile, disposable, one-patient… 02/02/2018Completed
Z-1336-2019 Class IITissu-Trans Syringe Fill 360, Catalog 3-TT-SFILL 360, sterile, disposable, one-patient…The pouch integrity of the sterile product can't be assured without a visual examination due to possibly… 02/02/2018Completed