Recall Z-1336-2010
Hospira Inc. · initiated November 24, 2009
Product
Lifeshield, CVP Manometer, latex-free, Y-type primary IV set, nonvented, 86.5 inch with pre-pierced injection site, and OPTION-LOK, 15 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made in Costa Rica, list 12729-65.
Codes / lots: Lot number: 600655H.
Reason for recall
Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.
FDA-determined cause: Component design/selection
Action taken
Hospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with the Borla designed CAIR clamp. The accounts were requested to quarantine all inventory of the affected lots, complete the enclosed reply form indicating the amount of product being returned, fax a copy of the completed form to Stericycle at 1-877-377-5114, and include a copy with the returned goods to Stericycle at 2670 Executive Drive, Suite A, Indianapolis, IN 46241.
Details
- Recall number
- Z-1336-2010
- Initiated
- November 24, 2009
- Posted by FDA
- April 15, 2010
- Terminated
- January 29, 2012
- Product code
- FPA – Set, Administration, Intravascular
- Quantity
- 2,832 sets
- Distribution
- Worldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Ireland, Italy, Jordan, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, San Marino…
- Recalling firm
- Hospira Inc., Lake Forest, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1327-2010 | Lifeshield, latex-free 100mL Burette Set, convertible pin, 77 inch with 2 pre-pierced…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1328-2010 | Lifeshield, latex-free 150 mL Burette IV Set, yellow key slide clamp, microdrip, CLAVE…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1329-2010 | Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 100 inch with 2 pre-pierced…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1330-2010 | Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 77 inch with capped port…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1331-2010 | Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 77.5 inch with 2…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1332-2010 | Lifeshield, latex-free, 150 mL Burette Set, nonvented, 117 inch with pre-pierced…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1333-2010 | Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 77 inch with 2 CLAVE ports…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1334-2010 | Lifeshield, latex-free 170 Micron Filter, Y-Type Blood Set, yellow key slide clamp…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1335-2010 | Lifeshield, latex-free 210 Micron Filter, Y-Type Blood Set, yellow key slide clamp…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1337-2010 | Lifeshield, latex-free, MICRO-DRIP, SOLUSET, GEMSTAR 150 ml Burette Pump Set-OL…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1338-2010 | Lifeshield, latex, free, GEMSTAR Primary Piggyback Pump Set-SL, convertible pin…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1339-2010 | Lifeshield, latex-free, HEMA, GEMSTAR, Y-Type Blood Pump Set, nonvented, 110 inch with…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1340-2010 | Latex-free, Nitroglycerin, Midlength Secondary I.V. Pump Set, convertible pin, 40 inch…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1341-2010 | Latex-free. Midlength Secondary I.V. Set, convertible pin, 40 inch with extension hook…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1342-2010 | ColorGard, latex-free, Midlength Secondary I.V. Set, convertible pin, 40 inch with…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1343-2010 | ColorGard, latex-free, Midlength Secondary I.V. Set, convertible pin, 40 inch with Green…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1344-2010 | ColorGard, latex-free, Midlength Secondary I.V. Set, convertible pin, 40 inch with…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1345-2010 | Latex-free, Nitroglycerin Primary I.V. Pump Set, convertible pin, 107 inch with…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1346-2010 | Lifeshield, Primary I.V. Filter Set, yellow key slide clamp, backcheck valve, 2 CLAVE…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1347-2010 | Lifeshield, Primary I.V. Filter Set, yellow key slide clamp, backcheck valve, 2 CLAVE…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1348-2010 | Lifeshield, latex-free, Primary I.V. PLUMSET, convertible pin, 113 inch with 2…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1349-2010 | Lifeshield, latex-free, Primary I.V. PLUMSET, convertible pin, 103 inch with backcheck…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1350-2010 | Latex-free, Primary I.V. Pump Set, convertible pin, 105 inch with 2 injection sites, 15…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1351-2010 | Latex-free, Primary I.V. Pump Set with distal microbore patient line and gravity flow…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1352-2010 | Lifeshield, latex-free, Primary I.V. Pump Set with distal microbore patient line and…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1353-2010 | Lifeshield, latex-free, Primary I.V. Pump Set with Distal Microbore Patient Line…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1354-2010 | Lifeshield, latex-free, Primary I.V. Pump Set with Distal Microbore Patient Line and…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1355-2010 | Lifeshield, latex-free, Primary I.V. Pump Set with Distal Microbore Patient Line and…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1356-2010 | Lifeshield, latex-free Primary I.V. Pump set with Distal Microbore Patient Line and…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |
| Z-1357-2010 | Lifeshield, latex-free, Primary I.V. Pump Set with Distal Microbore Patient Line…Reports of difficulty in adjusting the flow rate, which could result in flow rate variances. | 11/24/2009Terminated |