Recall Z-1336-2022

Chromsystems Instruments & Chemicals GmbH · initiated May 19, 2022

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level I

Codes / lots: Model: 0192; Lot Number: 2821; UDI/DI: 04250317502841

Reason for recall

Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products.

FDA-determined cause: Labeling Change Control

Action taken

The firm issued a recall notification on 05/19/2022 via email. The notification mentions if the consignee has used the affected product, check whether they have determined glycine, replace the old package insert with the updated package insert with release date 18.05.2022, forward a copy of the letter to another laboratory if the affected product was given to them, and submit the response form by 01.06.2022. If users determined glycine, they are asked to clarify with the attending physician or laboratory director, if necessary, whether a back-check of the values for glycine is required.

Details

Recall number
Z-1336-2022
Initiated
May 19, 2022
Terminated
May 15, 2024
Product code
JJY – Multi-Analyte Controls, All Kinds (Assayed)
Quantity
1330 units
Distribution
US: NY, OH, CT.
Recalling firm
Chromsystems Instruments & Chemicals GmbH, Grafelfing

Official FDA recall record