Recall Z-1339-2023

Stryker Corporation · initiated March 1, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,1PRT QUICK Catalog number: 5921-030-135NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities

Codes / lots: GTIN: 07613327606157 Lot Numbers; 2022111803 2022112901 2022121911 2022121912 2022122601 2022122602

Reason for recall

Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications

FDA-determined cause: Under Investigation by firm

Action taken

Stryker issued Urgent Medical Device Recall letter dated 2/27/23 via 2-day Fedex delivery. The purpose of this letter is to inform you of an increase in reported low pressure events for specific lots of Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuffs and to reiterate the steps in the Instructions For Use (IFU) to mitigate the risk to achieve or maintain pressure. Letter states reason for recall, health risk and action to take: Please review all action steps below and complete the acknowledgement section (pg.3) of this letter and return by email to Instruments.recalls@stryker.com. 1. Immediately review this notification and distribute this letter internally to all users of the Color Cuff Non-Sterile Disposable Tourniquet Cuffs. 2. Follow the subsequent steps from the Instructions For Use (IFU) to mitigate the risk of the cuff not being able to achieve or maintain pressure. Following these steps, you may continue to use the product as-is. a. Prior to using the Non-Sterile Tourniquet Cuffs, please review the Safety Directives, Inspection, and Instructions sections of the IFU. To view this IFU electronically, go to https://ifu.stryker.com/XHR/129532. b. Prior to connecting the Non-Sterile Disposable Tourniquet Cuffs to the Stryker SmartPump, please review the bullet points below from pages 18-20 of the IFU for the SmartPump Tourniquet System. For troubleshooting, please refer to page 28. To view this IFU electronically, go to https://ifu.stryker.com/XHR/129013. " ALWAYS monitor pneumatic connections during use, including the cuff(s) and cuff connector(s) with O-ring, the quick-lock connectors, and the fill line(s) and connectors. If leakage or damage is apparent, see Troubleshooting. " NOTE: If a cuff does not reach or maintain the target pressure value during initial inflation, discontinue using the cuff. See To Switch (or Backup) a Single Cuff. " ALWAYS respond to an alert condition before continuing to use the tourniquet pump. An ALERT symbol, a flashing

Details

Recall number
Z-1339-2023
Initiated
March 1, 2023
Posted by FDA
April 4, 2023
Product code
KCY – Tourniquet, Pneumatic
Quantity
1225 units (packs of 10)
Distribution
US Nationwide distribution.
Recalling firm
Stryker Corporation, Portage, MI

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1340-2023 Class IIStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 30X4,1BLA,2PRT QUICK…Not able to achieve desired pressure and not holding pressure result from two different causes which have… 03/01/2023Open, Classified
Z-1341-2023 Class IIStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,1PRT QUICK…Not able to achieve desired pressure and not holding pressure result from two different causes which have… 03/01/2023Open, Classified
Z-1342-2023 Class IIStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,2PRT QUICK…Not able to achieve desired pressure and not holding pressure result from two different causes which have… 03/01/2023Open, Classified
Z-1343-2023 Class IIStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 18X4,1 BLA,1PRT QUICK…Not able to achieve desired pressure and not holding pressure result from two different causes which have… 03/01/2023Open, Classified
Z-1344-2023 Class IIStryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 18X4,1BLA,2PRT QUICK…Not able to achieve desired pressure and not holding pressure result from two different causes which have… 03/01/2023Open, Classified