Recall Z-1352-2015

Synthes, Inc. · initiated February 24, 2015

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures in adults and adolescents (12-21) in which the growth plates have fused.

Codes / lots: Part Numbers: 03.037.013, 03.037.014, Lot Numbers: 8903848, 9124017, 903851, 9124019, 8918566, 9124022, 9124012, 9124023, 9124014, 9124026, 9124016, 9124027, 8918563

Reason for recall

The guide sleeve has the potential to jam in the Aiming Arms, which may delay disassembly and removal of instruments.

FDA-determined cause: Component design/selection

Action taken

DePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated March 2, 2015, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to review inventory for the affected parts and following the actions provided in the letter for removal and replacement of these devices. For questions call 610-719-5450.

Details

Recall number
Z-1352-2015
Initiated
February 24, 2015
Posted by FDA
March 30, 2015
Terminated
September 21, 2015
Product code
HSB – Rod, Fixation, Intramedullary And Accessories
Quantity
131
Distribution
US Nationwide Distribution - NY, TX, MD, OR, MS, ND, CA, AZ, MI, IL, PA, TN, AR, SD, SC, FL, VA, OK, and GA
Recalling firm
Synthes, Inc., West Chester, PA

Official FDA recall record