Recall Z-1354-2026
Olympus Corporation of the Americas · initiated January 16, 2026
Class I – dangerous or defective products that could cause serious health problems or death
Product
Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.
Codes / lots: Model Number: UHI-3; UDI-DI: 04953170140280, 04953170140297; All Serial Numbers.
Reason for recall
Issue with software algorithm which may lead to overpressure events.
FDA-determined cause: Software design
Action taken
Olympus notified consignees on about 01/16/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to cease usage of the affected product immediately. They are advised to use available alternatives including an existing UHI-4 if available or switching to an insufflation device from another manufacturer. If no alternative is available on site, contact Olympus to arrange a loaner or purchase a replacement. Additionally, consignees were instructed to examine inventory and quarantine any identified devices, immediately cease usage of any UHI, UHI-2, and/or UHI-3 in inventory, call Olympus Customer Service at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return your device. Consignees were also requested to acknowledge receipt of this letter through their recall web portal as follows: a. Go to https://olympusamerica.com/recall, b. Enter the recall number "0489 and c. Complete the form as instructed. Please indicate in your response if you no longer have the device(s). If any consignees have further distributed this UHI product, they were instructed to identify affected customers and forward this notice to them immediately. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by email at Cynthia.Ow@Olympus.com.
Details
- Recall number
- Z-1354-2026
- Initiated
- January 16, 2026
- Posted by FDA
- February 18, 2026
- Product code
- HIF – Insufflator, Laparoscopic
- Quantity
- 485 units
- Distribution
- Nationwide distribution
- Recalling firm
- Olympus Corporation of the Americas, Center Valley, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1352-2026 Class I | Olympus High Flow Insufflation Unit. Model Number: UHI. Used for insufflation of the…Issue with software algorithm which may lead to overpressure events. | 01/16/2026Open, Classified |
| Z-1353-2026 Class I | Olympus High Flow Insufflation Unit. Model/ Number: UHI-2. Used for insufflation of the…Issue with software algorithm which may lead to overpressure events. | 01/16/2026Open, Classified |