Recall Z-1363-2016

Natus Medical Incorporated · initiated January 8, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator.

Codes / lots: Serial numbers: 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5294; 5293; 5292; 5291; 5290; 5289; 5287; 5283; 5282; 5281; 5280; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; 5129; 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; 5045. Serial numbers were updated in the final status report. They are as follows: 5552; 5551; 5550; 5549; 5510; 5509; 5508;5507; 5506; 5505; 5504; 5389; 5388; 5385; 5383; 5337; 5336; 5335; 5334; 5333; 5332; 5331; 5330; 5329; 5328; 5327; 5326; 5325; 5324; 5323; 5322; 5294; 5293; 5292; 5291; 5290; 5289; 5288; 5287; 5286; 5285; 5284; 5283; 5282; 5281; 5280; 5279; 5278; 5277; 5276; 5257; 5256; 5255; 5254; 5253; 5252; 5251; 5250; 5249; 5248; 5228; 5227; 5226; 5225; 5224; 5223; 5222; 5221; 5220; 5219; 5214; 5213; 5212; 5211; 5210; 5209; 5139; 5138; 5136; 5135; 5134; 5133; 5132; 5131; 5130; and 5129. The following serial numbers were removed: 5097; 5096; 5057; 5056; 5053; 5051; 5048; 5047; 5046; and 5045.

Reason for recall

NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator.

FDA-determined cause: Process control

Action taken

Natus of Seattle, WA began sending the Urgent Field Safety Notice NatalCare LX Incubator letter (DOC-017939A) to consignees on 1/8/16. A repair kit to correct this issue is now available for repair of affected units. Consignees should complete the NatalCare LX Incubator Oval Port Hinge Repair Kit Order Form and FAX it 847-949-6031 or scan and email this form to Natus_Quality_Programs@natus.com. Alternately you can telephone Natus Technical Service at 888-496-2887 and order the kit from a specialist. Natus will send the kit(s) free-of-charge to the address you specify. Each upgrade will require up to - 1 hour of time and require simple hand tools. If you have any questions, please contact Technical Service at 888-496-2887 or send email to: customer.service@natus.com

Details

Recall number
Z-1363-2016
Initiated
January 8, 2016
Posted by FDA
April 11, 2016
Terminated
February 10, 2017
Product code
FMZ – Incubator, Neonatal
Quantity
75 units (65 units in the US and 10 units in Pakistan) updated in final status report 36 customers with 96 units.
Distribution
Distribution US nationwide and Pakistan.
Recalling firm
Natus Medical Incorporated, Seattle, WA

Official FDA recall record