Recall Z-1364-2019

Ortho-Clinical Diagnostics · initiated March 20, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest

Codes / lots: Serial Number 76000108 thru current.

Reason for recall

Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported

FDA-determined cause: Software design

Action taken

Ortho Clinical Diagnostics issued on 20 March 2019, a customer letter via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) customers who were shipped affected VITROS XT 7600 Systems. Software Version 3.4.2 (MOD 4) will be provided for download or DVD format. Contact Ortho Care" Technical Solutions Center at 1-800-421-3311.

Details

Recall number
Z-1364-2019
Initiated
March 20, 2019
Terminated
December 8, 2020
Product code
JJE – Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Quantity
85 units: US=19;OUS=66
Distribution
Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY

Official FDA recall record