Recall Z-1371-2013
Nephron Pharmaceuticals Corp. · initiated April 30, 2013
Class I – dangerous or defective products that could cause serious health problems or death
Product
EZ Breathe Atomizer Model # EZ-100. For the delivery of liquid medications for respiratory usage.
Codes / lots: Model # EZ-100 Lots # R2029A, R2029B, R2039A, R2042A, R2045A, and R2047A. Atomizer Serial Number Ranges: 1206034476-1206069065 1207003710-1237038299 1207046505-1207081124 1208027421-1208062155 1209000001-1209069179 1209069180-1209069202 1209069203-1209069460 1210000001-1210103680 1210104001-1210104044
Reason for recall
Possible dislodgement of the "Plate A", if this instance occurs, it presents a potential choking hazard.
FDA-determined cause: Nonconforming Material/Component
Action taken
The firm, Nephron Pharmaceuticals Corporation, notified its consignees on April 30, 2013, and May 3, 2013, of the problem via a recall notice and a Press Release, respectively. The firm directed consignees to discontinue use of any EZ Breathe atomizer units contained in the Asthmanefrin Starter Kits with the affected lot numbers and also those that were sold individually. Customers were directed to return all affected product.
Details
- Recall number
- Z-1371-2013
- Initiated
- April 30, 2013
- Posted by FDA
- May 29, 2013
- Terminated
- June 28, 2016
- Product code
- CCQ – Nebulizer, Medicinal, Non-Ventilatory (Atomizer)
- Quantity
- 604 each
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AZ, CA, CO, CT, FL, IL, IN, IO, GA, KS, KY, LA, MD, MN, MS, MO, NK, NV, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN , TX, VT, VA, WA, and WV, and the countries of Guatemala, Poland, Panama, and United Kingdom.
- Recalling firm
- Nephron Pharmaceuticals Corp., Orlando, FL