Recall Z-1389-2025
Immuno-Mycologics, Inc · initiated February 10, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE
Codes / lots: Lot #: F4171005/UDI: 00816387020223
Reason for recall
Candida Immunodiffusion Antigen may have reduced sensitivity which could result in false negative results.
FDA-determined cause: Process change control
Action taken
On February 10, 2025, Immy issued a "Urgent Medical Device Recall" Notification to affected consignees. Immy asked consignees to take the following actions: 1. Immediately identify, count, and segregate any affected vials you have in your inventory to prevent them from being used or shipped to your customers. As soon as possible, discard your affected inventory. 2. Complete the attached Acknowledgement and Receipt Form (pages 3 and 4 below) even if you do not have any affected stock remaining in your possession. Note: The form is a fillable PDF. You can save it to your computer, fill out electronically and attach to an email. Return the completed form to IMMY using one of the methods below: o Email: customerservice@immy.com; Mail to: Attn: Hunter Conover IMMY, Inc., 2701 Corporate Centre Dr., Norman, OK USA 73069 3. Ensure relevant staff members are informed of this recall, including relevant clinicians. Clinicians should review all negative results. 4. If you have supplied any potentially affected product to another organization, please advise that organization of this recall and send them this notification. Please contact us so we can follow up with them. 5. In case product is in transit, display this letter in a prominent place for one month.
Details
- Recall number
- Z-1389-2025
- Initiated
- February 10, 2025
- Posted by FDA
- March 18, 2025
- Product code
- GMI – Antigen, Cf And / Or Id, Coccidioides Immitis
- Quantity
- 30 vials
- Distribution
- Worldwide - US Nationwide distribution in the states of CA and OK and the countries of Peru France, Austria, Venezuela.
- Recalling firm
- Immuno-Mycologics, Inc, Norman, OK