Recall Z-1403-2025

Zimmer Orthopedics Manfacturing Limited-Galway · initiated March 3, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627

Codes / lots: Item No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215; Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219; Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221; Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809

Reason for recall

Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.

FDA-determined cause: Nonconforming Material/Component

Action taken

On March 3, 2025, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were informed of the excess metal on the inner taper of the stem due to a chipped manufacturing tool. Customers were asked to notify all affected persons and assist their Zimmer Biomet sales representative in quarantining all affected product on hand. The products will be removed from your facility.

Details

Recall number
Z-1403-2025
Initiated
March 3, 2025
Posted by FDA
March 20, 2025
Product code
MBF – Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented
Quantity
90 US; 136 OUS
Distribution
US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.
Recalling firm
Zimmer Orthopedics Manfacturing Limited-Galway, Oranmore

Official FDA recall record