Recall Z-1408-2023
Medtronic Neuromodulation · initiated March 17, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Medtronic SenSight Extension Tunneler Kit, REF B31030; The Extension Tunneler Kit is used to assist in the tunneling and passage of extensions for deep brain stimulation
Codes / lots: UDI/DI00763000426361, Serial Numbers: HG56BR6, HG56C4R, HG56CJAH01, HG56CLE, HG586N7, HG586UM, HG58704, HG58VM0, HG58VQH, HG58VQHH01, HG58VY5, HG58W7H, HG58WWC, HG58WYB, HG58WYBH01, HG58X0E, HG597BH, HG597DH, HG597M8, HG59ABH, HG59AFC, HG59AV6, HG59RY0, HG5A9GT, HG5A9VY, HG5A9VZ, HG5AQH0, HG5AQQE, HG5BQ5Y, HG5BQLR, HG5BQPJ, HG5BTW7, HG5BTZX, HG5BU5W, HG5BU77, HG5EPSB, HG5EQ0C, HG5EQ2A, HG5EQ69, HG5F31N, HG5F338, HG5F3AZ, HG5F3EH, HG5F3G6, HG5F3K5, HG5F3NF, HG5FP0Y, HG5FP3B, HG5FP70, HG5FP70H01, HG5FPBE, HG5HQBM, HG5HQF8, HG5LJTL, HG5LJZQ, HG5LK2J, HG5LK4H, HG5LK5Y, HG5LK7C, HG5LK8P, HG5LK92, HG5LKAL, HG5LKC9, HG5LKEL, HG5LKGT, HG5LR02, HG5LR33, HG5LTWA, HG5LTWP, HG5LTXU, HG5LTYY, HG5LU12; UDI/DI 00763000486518, Serial Numbers: HG59ABH, HG59RY0, HG59SJ4, HG59SNC, HG5A9H3, HG5A9NB, HG5A9VY, HG5BQ5Y, HG5BQAW, HG5BQLR.
Reason for recall
Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.
FDA-determined cause: Nonconforming Material/Component
Action taken
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/17/2023 via UPS 2-day delivery. The notice explained the issue and the hazard and requested the following actions be taken: Quarantine all unused product from the affected lots and return them as described on the Customer Confirmation Form. Pass the notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed.
Details
- Recall number
- Z-1408-2023
- Initiated
- March 17, 2023
- Posted by FDA
- April 19, 2023
- Product code
- MHY – Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Quantity
- 1414 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Portugal, Slovakia, Spain, Sweden, Switzerland, United Kingdom.
- Recalling firm
- Medtronic Neuromodulation, Minneapolis, MN