Recall Z-1416-2009

Cook OB/Gyn, Inc. · initiated April 2, 2009

Terminated

Product

Cook Medical Tao Brush I.U.M.C. Endometrial Cytology/Histology Sampler (Tao Brush), 9.0Fr/26 cm, 3.5 cm, sterile, Cook Ob/Gyn, Spencer, IN; REF J-ES-090500-BL. Used to obtain endometrial samples for both cytology and histology. Intended for one-time use.

Codes / lots: Lots UF1793872 through UF1793877, UF1811910 through UF1811913, UF1811916, UF1811917, UF1811917S, UF1811918, UF1811919, UF1816687 and UF1816688.

Reason for recall

There is an ink on the brush which has not been tested for use with the Tao Brush.

FDA-determined cause: Process control

Action taken

The consignee was notified by an Urgent: Recall Notice letter dated 4/2/09. The customer is to forward the notice to all personnel that need to be aware of it; pull from their stock and isolate any affected product; and complete the form. Upon receipt of the form, Cook Women's Health will issue a Return Goods Authorization number and the customer should return the affected products via FedEx using the Return Goods Authorization number.

Details

Recall number
Z-1416-2009
Initiated
April 2, 2009
Posted by FDA
June 16, 2009
Terminated
December 3, 2009
Product code
HFE – Brush, Endometrial
Quantity
15,396
Distribution
Nationwide Distribution.
Recalling firm
Cook OB/Gyn, Inc., Spencer, IN