Recall Z-1416-2018

Terumo Cardiovascular Systems Corporation · initiated March 16, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Terumo HX2 Temperature Management System, Catalog No. 809810 The Terumo HX2 Temperature Management System is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.

Codes / lots: All Lot Codes

Reason for recall

Update to cooler-heater cleaning instructions.

FDA-determined cause: Labeling design

Action taken

The firm,Terumo Cardiovascular Systems Corporation, mailed customers an "Urgent Medical Device Correction" letter dated 2/16/18. The letter described the product, problem and actions to be taken. The Customers were instructed to do the following: 1) discard the current Sarns TCM, TCM II Cooling and Heating, or tHX2 Temperature Management Systems cleaning guides; 2) Review the notice and updated cleaning guides, and ensure all users receive notice of the updated cleaning guides and; 3) Return the Customer Response Form acknowledging receipt of the notice via Terumo Recall Email: tcvs.recall@terumomedical.com or Terumo Recall Fax: 734-741-6149. The letter also includes a list of the updates to the cleaning guides. Customers with questions may contact Terumo CVS Customer Service (800) 521-2818, Monday-Friday, 8am-6pm ET or Fax (734)741-6149.

Details

Recall number
Z-1416-2018
Initiated
March 16, 2018
Posted by FDA
April 18, 2018
Terminated
June 10, 2020
Product code
DWC – Controller, Temperature, Cardiopulmonary Bypass
Quantity
2393 this total is for all products
Distribution
Worldwide Distribution: US (Nationwide), Australia, Belgium, Canada, Chile, China. Colombia, Dominican Republic, England, Finland, France, Germany, Greece, HK, India, Indonesia, Iran. Israel, Italy, Japan, Korea, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates,and…
Recalling firm
Terumo Cardiovascular Systems Corporation, Ann Arbor, MI

Official FDA recall record