Recall Z-1422-2020

Beckman Coulter Inc · initiated August 15, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Enzymatic Creatinine Reagent (CR-E 2 X 200), REF A60298, For in Vitro Diagnostic Use, Rx Only, UDI: 15099590625733

Codes / lots: All Lot Numbers

Reason for recall

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.

FDA-determined cause: Device Design

Action taken

On August 15, 2019, the firm mailed or emailed a "Urgent Medical device Recall" to affected consignees In addition to informing the consignees about the recall, they ask consignees to do the following: 1. Resolution: The following statement will be added to the Interfering Substances section of the Enzymatic Creatinine, Triglycerides GPO Blanked, Uric Acid, Direct Bilirubin, and Total Bilirubin Chemistry Information Sheets (CIS): N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may generate erroneously low results in samples for patients that have taken toxic doses of acetaminophen (paracetamol). 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 3. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email. " Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center; " From our website: http://www.beckmancoulter.com " By phone: call 1-800-854-3633 in the United States and Canada. " Outside the United States and Canada, contact your local Beckman Coulter representative.

Details

Recall number
Z-1422-2020
Initiated
August 15, 2019
Terminated
February 28, 2024
Product code
JFY – Enzymatic Method, Creatinine
Quantity
2581 Reagents
Distribution
Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South…
Recalling firm
Beckman Coulter Inc, Carlsbad, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1423-2020 Class IITriglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitro Diagnostic Use, Rx Only…It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen… 08/15/2019Terminated
Z-1424-2020 Class IIUric Acid Reagent (URIC 2 X 300), REF: 442785, For In Vitro Diagnostic Use, Rx Only…It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen… 08/15/2019Terminated
Z-1425-2020 Class IIDirect Bilirubin Reagents (DBIL 2 x 200 ) and (DBIL 2 x 300), Ref:439715 (200…It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen… 08/15/2019Terminated
Z-1426-2020 Class IITotal Bilirubin Reagent (TBIL 2 x 300 ) and (TBIL 2 x 400), Ref:442745 (300…It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen… 08/15/2019Terminated