Recall Z-1423-2022

Randox Laboratories Ltd. · initiated June 9, 2022

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD

Codes / lots: UDI-DI (GTIN): 05055273215557 All lots within expiry

Reason for recall

A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.

FDA-determined cause: Under Investigation by firm

Action taken

On June 9, 2022, the firm notified customers of the recall via email. The issue only affects Siemens Atellica CH analyzer users. No action is required if the assay is being run on other platforms. The issue will be resolved in Atellica Solution Software (SW) version 1.25.3. Until all Atellica CH analyzers in the laboratory are updated to SW 1.25.3, the assay can be run by Open Channel. Please contact applications@randox.com for an Atellica CH Open Channel application sheet. If the assay is not run by Open Channel prior to installation of SW version 1.25.3, the assay must be re-calibrated once 1.25.3 is installed. Customers were instructed to review results generated with the affected batches in line with the clinical profile of the patient.

Details

Recall number
Z-1423-2022
Initiated
June 9, 2022
Terminated
February 8, 2024
Product code
DDG – Transferrin, Antigen, Antiserum, Control
Quantity
12 kits (US)
Distribution
US Nationwide distribution in the state of IL.
Recalling firm
Randox Laboratories Ltd., Crumlin, North

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1422-2022 Class IIISoluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVDA customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an… 06/09/2022Terminated