Recall Z-1425-2012

Abbott Laboratories Inc. · initiated July 9, 2009

Terminated

Product

Clinical Chemistry Total Bilirubin, 6L45-20 and 6L45-40. Used for the quantitative analysis of total bilirubin in human serum or plasma of adults and neonates.

Codes / lots: List Numbers: 6L45-20, 6L45-40 Lot Numbers: 64047HW00, 64048HW00, 65084HW00, 65088HW00, 69001HW00, 69002HW00, 72016HW00, 73038HW00, 73057HW00, 75036HW00, 76023HW00, 77014HW00, 69025HW00, 73015HW00, 76038HW00, 77015HW00.

Reason for recall

Abbott Laboratories has issued a recall on the Clinical Chemistry Total Bilirubin due to an under-recovery on one clinical neonatal specimen resulting in a value less than the assay linear limit.

FDA-determined cause: Material/Component Contamination

Action taken

Abbott Laboratories plans to release a Product Correction letter to customers instructing customers to follow the steps provided in order to revise their Total Bilirubin assay parameters. The letter explained the reason for the recall, along with the patient impact. Customers were to keep the letter as part of their laboratory records and also forward a copy to any other laboratories that they may have forwarded Total Bilirubin kits. U.S. Customers should contact Customer Service at 1-877-4ABBOTT, if they have questions.

Details

Recall number
Z-1425-2012
Initiated
July 9, 2009
Posted by FDA
April 16, 2012
Terminated
March 7, 2014
Product code
CIG – Diazo Colorimetry, Bilirubin
Quantity
11,282
Distribution
Worldwide Distribution -- USA, Guatemala, Chile, Brazil, Canada, Germany, Hong Kong, Thailand, Puerto Rico, Australia, New Zealand, Trinidad & Tobago, Bermuda, Curacao, Cayman Island, and Jamaica.
Recalling firm
Abbott Laboratories Inc., South Pasadena, CA

Official FDA recall record