Recall Z-1429-2025
MEDLINE INDUSTRIES, LP - Northfield · initiated February 12, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
namic convenience kits labeled as: KIT,ARTERIAL,60,(152CM),PG; medical convenience kit, REF 70036160
Codes / lots: UDI/DI 10193489064803 (each), 20193489064800 (case), Lot number 0000140054
Reason for recall
The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
FDA-determined cause: Process change control
Action taken
Medline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medline will provide labeling that states the following: "ATTENTION This kit contains a recalled item Medline Pressure Monitoring Lines VPN: 24614303 DO NOT USE THIS COMPONENT The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*" Consignees were directed to notify their customers of these actions.
Details
- Recall number
- Z-1429-2025
- Initiated
- February 12, 2025
- Posted by FDA
- March 21, 2025
- Product code
- DRS – Transducer, Blood-Pressure, Extravascular
- Quantity
- 40 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1428-2025 Class II | namic convenience kits labeled as: PML,72,ADULT,FLL-MLL,-,OEM, REF 80641722; medical…The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near… | 02/12/2025Open, Classified |
| Z-1430-2025 Class II | namic convenience kits labeled as: MTS,COLOR,PML,KIT,-,FRENCH HOSPITAL PG; medical…The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near… | 02/12/2025Open, Classified |
| Z-1431-2025 Class II | namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468…The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near… | 02/12/2025Open, Classified |