Recall Z-1429-2025

MEDLINE INDUSTRIES, LP - Northfield · initiated February 12, 2025

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

namic convenience kits labeled as: KIT,ARTERIAL,60,(152CM),PG; medical convenience kit, REF 70036160

Codes / lots: UDI/DI 10193489064803 (each), 20193489064800 (case), Lot number 0000140054

Reason for recall

The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.

FDA-determined cause: Process change control

Action taken

Medline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medline will provide labeling that states the following: "ATTENTION This kit contains a recalled item Medline Pressure Monitoring Lines VPN: 24614303 DO NOT USE THIS COMPONENT The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*" Consignees were directed to notify their customers of these actions.

Details

Recall number
Z-1429-2025
Initiated
February 12, 2025
Posted by FDA
March 21, 2025
Product code
DRS – Transducer, Blood-Pressure, Extravascular
Quantity
40 units
Distribution
Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL

Official FDA recall record