Recall Z-1431-2020

Bard Peripheral Vascular Inc · initiated January 3, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Bard Biopsy EnCor Probe, Stereotactic/Ultrasound 7G, REF number ECP017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.

Codes / lots: Lot numbers VTCT0268, VTCT0269, VTCT0270, VTCT0271, VTCU0340, and VTCU0341. UDI numbers: (01)00801741086281(17)200601(10)VTCT0268, (01)00801741086281(17)200601(10)VTCT0269, (01)00801741086281(17)200601(10)VTCT0270, (01)00801741086281(17)200601(10)VTCT0271, (01)00801741086281(17)200601(10)VTCU0340, and (01)00801741086281(17)200601(10)VTCU0341.

Reason for recall

Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.

FDA-determined cause: Component change control

Action taken

The recalling firm issued letters dated 1/3/2020 to each of the U.S. consignees via FedEx with proof of delivery notification. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-321-4254 Option #5.

Details

Recall number
Z-1431-2020
Initiated
January 3, 2020
Terminated
March 17, 2022
Product code
KNW – Instrument, Biopsy
Quantity
12,875 units
Distribution
Distribution was made to AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, SC, TX, VA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Canada, Australia, Austria, Belgium, China, Czech Republic, Germany, France, Georgia, Greece, Hong Kong, Israel, Italy, Japan, Korea, Netherlands, New…
Recalling firm
Bard Peripheral Vascular Inc, Tempe, AZ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1432-2020 Class IIBard Biopsy EnCor Probe, MRI, 7G Trocar Tip, REF number ECPMR017G, packaged individually…Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was… 01/03/2020Terminated
Z-1433-2020 Class IIBard Biopsy EnCor Probe, MRI, 10G Trocar Tip, REF number ECPMR0110G, packaged…Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was… 01/03/2020Terminated
Z-1434-2020 Class IIBard Biopsy EnCor Probe, MRI, 10G Blunt Tip, REF number ECPMR0110GBT, packaged…Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was… 01/03/2020Terminated