Recall Z-1436-2014

Clinical Diagnostic Solutions · initiated February 24, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Boule Con-Diff Tri-Level Multi-Parameter Assayed Hematology Control, used for in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments.

Codes / lots: Product Part Number: 501-605 Lot# 1311-682 Exp. 3/26/2014 501-607 Lot# 1311-683, 1311-684 Exp. 3/26/2014 501-608 Lot# 1311-683 Exp. 3/26/2014

Reason for recall

Elevated MCV results on the 1311-682, 1311-683, and 1311-684 lots.

FDA-determined cause: Under Investigation by firm

Action taken

Clinical Diagnostic Solutions sent a Notification letter dated February 26, 2014 to affected customers. The letter identified the affected product, problem and the actions to be taken. For questions contact Technical Support at 1-800-453-3328.

Details

Recall number
Z-1436-2014
Initiated
February 24, 2013
Posted by FDA
April 9, 2014
Terminated
August 7, 2014
Product code
JPK – Mixture, Hematology Quality Control
Quantity
3,540
Distribution
Worldwide Distribution: USA Nationwide and the countries of IVORY COAST, MEXICO, and Philippines.
Recalling firm
Clinical Diagnostic Solutions, Plantation, FL