Recall Z-1436-2025
MEDLINE INDUSTRIES, LP - Northfield · initiated February 11, 2025
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Medline procedure kits labeled as: 1) ANGIO PACK, Pack Number DYNJ50624A; 2) ANGIO PACK, Pack Number DYNJ30501F; 3) ANGIO PACK-LF, Pack Number DYNJ0954970G; 4) ANGIOGRAPHY 6 FRENCH PACK, Pack Number DYNJ33597J; 5) ANGIOGRAPHY DRAPE PACK, Pack Number DYNJ38120B; 6) ANGIOGRAPHY DRAPE PACK, Pack Number DYNJ42873; 7) ANGIOGRAPHY PACK, Pack Number DYNJ57498; 8) ANGIOGRAPHY PACK, Pack Number DYNJ52167A; 9) ANGIOGRAPHY PACK, Pack Number DYNJV0173I; 10) ANGIOGRAPHY TRAY-LF, Pack Number DYNJ58722A; 11) CATH LAB PACK, Pack Number DYNJ42912F; 12) CATH LAB PK, Pack Number DYNJ35643B; 13) CATH PACK-LF, Pack Number DYNJ0367643Q; 14) CRMS ANGIO PACK-LF, Pack Number PHS616026002A; 15) DEPAUL ANGIO CATH PACK, Pack Number DYNJ24542L
Codes / lots: 1) Pack Number DYNJ50624A: UDI/DI 10889942474639 (each) 40889942474630 (case), Lot Number 20BDB393A; 2) Pack Number DYNJ30501F: UDI/DI 10889942806683 (each) 40889942806684 (case), Lot Number 21HBP134A; 3) Pack Number DYNJ0954970G: UDI/DI 10193489505634 (each) 40193489505635 (case), Lot Number 21GMA850A; 4) Pack Number DYNJ33597J: UDI/DI 10889942802012 (each) 40889942802013 (case), Lot Number 21IMD296A; 5) Pack Number DYNJ38120B: UDI/DI 10884389722145 (each) 40884389722146 (case), Lot Number 21BBI114A; 6) Pack Number DYNJ42873: UDI/DI 10888277011298 (each) 40888277011299 (case), Lot Number 23EBE560A; 7) Pack Number DYNJ57498: UDI/DI 10889942745746 (each) 40889942745747 (case), Lot Number 21AMA569A; 8) Pack Number DYNJ52167A: UDI/DI 10193489541106 (each) 40193489541107 (case), Lot Number 21GMB606A; 9) Pack Number DYNJV0173I: UDI/DI 10889942965021 (each) 40889942965022 (case), Lot Number 23CLA070A; 10) Pack Number DYNJ58722A: UDI/DI 10193489329698 (each) 40193489329699 (case), Lot Number 21IME955A; 11) Pack Number DYNJ42912F: UDI/DI 10193489785753 (each) 40193489785754 (case), Lot Number 23LBU921Z; 12) Pack Number DYNJ35643B: UDI/DI 10193489223712 (each) 40193489223713 (case), Lot…
Reason for recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
FDA-determined cause: Under Investigation by firm
Action taken
Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. This is an expansion of RES 94583
Details
- Recall number
- Z-1436-2025
- Initiated
- February 11, 2025
- Posted by FDA
- March 25, 2025
- Product code
- OEQ – Angiography/Angioplasty Kit
- Quantity
- 1939 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1435-2025 Class II | Medline procedure kits labeled as: 1) BILATERAL FACET TRAY, Pack Number DYNJRA0869A; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1437-2025 Class II | Medline procedure kits labeled as: 1) ANGIO DRAPE PACK (DRLUL)-LF, Pack Number…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1438-2025 Class II | Medline procedure kits labeled as: 1) AV FISTULA PACK, Pack Number DYNJ80088; 2) AV…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1439-2025 Class II | Medline procedure kits labeled as: 1) AV FISTULA PACK-LF, Pack Number DYNJ0826753O; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1440-2025 Class II | Medline procedure kits labeled as: 1) CT BIOPSY PACK, Pack Number DYNJ62734A; 2) E R C P…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1441-2025 Class II | Medline procedure kits labeled as: 1) ABLATION PACK, Pack Number DYNJ40629B; 2) ACH…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1442-2025 Class II | Medline procedure kits labeled as: 1) GENERAL ENDOSCOPY PACK, Pack Number DYNJ44673GMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1443-2025 Class II | Medline procedure kits labeled as: 1) CV NEONATAL PACK, Pack Number DYNJ67356Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1444-2025 Class II | Medline procedure kits labeled as: 1) A & A KNEE PACK-LF, Pack Number DYNJHS0136D; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1445-2025 Class II | Medline procedure kits labeled as: 1) LACERATION TRAY, Pack Number DYNJ17328BMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1446-2025 Class II | Medline procedure kits labeled as: 1) DELIVERY PACK, Pack Number DYNJ0923016; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1447-2025 Class II | Medline procedure kits labeled as: 1) CLOVIS ROBOTIC-LF, Pack Number DYNJ61061D; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1448-2025 Class II | Medline procedure kits labeled as: 1) C SECTION, Pack Number DYNJ57622; 2) C SECTION…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1449-2025 Class II | Medline procedure kits labeled as: 1) CSMC/ABDOMINAL HYST PACK-LF, Pack Number…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1450-2025 Class II | Medline procedure kits labeled as: 1) L S GYN PACK, Pack Number DYNJ65616C; 2) MAIN…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1451-2025 Class II | Medline procedure kits labeled as: 1) AOP/BASIC EYE PACK-LF, Pack Number DYNJ0589795V…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1452-2025 Class II | Medline procedure kits labeled as: 1) ARTHROSCOPY PACK, Pack Number DYNJHS0101I; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |
| Z-1453-2025 Class II | Medline procedure kits labeled as: 1) C SCOPE DISPOSABLE KIT, Pack Number 613034700, 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… | 02/11/2025Open, Classified |