Recall Z-1437-2011
Gen-Probe Inc · initiated December 30, 2008
Product
Mycobacterium Avium Complex Culture Identification Test Identification Test Kit, 2835.
Codes / lots: Lots (batches): 555767
Reason for recall
This recall has been initiated following the identification of certain assay tubes that do not contain probe reagent. The affected kits may contain a combination of tubes with correct probe amount, partial probe amount and some probe tubes that are empty. Possible false negative results.
FDA-determined cause: Process control
Action taken
Gen-Probe issued a Recall Notice to customers recommending that they discontinue the use of the identified batch (lot) numbers listed. For customers that have used these batches and obtained negative results, they recommend that the referring physician be notified about the possibility of a false negative result. With the Recall Notice, consignees received a Customer Response Form to complete. Gen-Probe requested that they segregate the affected batches listed on the Customer Response Form, then reconcile, and destroy all partial and/or full kits left in inventory, being sure to document the amount destroyed on the Customer Response Form. Gen-Probe should be updated via phone, fax, and email information. Customers should then sign and return the completed form to Gen-Probe via fax to 858-410-8250, or scanned and emailed to technicalsupport@genprobe.com. The Technical Support team should be contacted for replacement kits. Technical Support was also available for additional information at 888-484-4747 or technicalsupport@gen-probe.com.
Details
- Recall number
- Z-1437-2011
- Initiated
- December 30, 2008
- Posted by FDA
- March 10, 2011
- Terminated
- June 8, 2011
- Product code
- LQF – Dna-Reagents, Mycobacterium Spp.
- Quantity
- 311 kits
- Distribution
- Worldwide Distribution -- USA, including Puerto Rico and countries of Australia, Guatemala, Malaysia, Bahamas, Hong Kong, Mali, Singapore, Bermuda, India, Mexico, South Africa, Canada, Israel, Netherlands, South Korea, Finland, Kuwait, Panama, Thailand, France, Latvia, and Philippines.
- Recalling firm
- Gen-Probe Inc, San Diego, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-1436-2011 | Gen-Probe AccuProbe Group B Streptococcus Culture Identification Test Kit, 2820.This recall has been initiated following the identification of certain assay tubes that do not contain probe… | 12/30/2008Terminated |
| Z-1438-2011 | Mycobacterium Tuberculosis Complex Culture Identification Test, 2860.This recall has been initiated following the identification of certain assay tubes that do not contain probe… | 12/30/2008Terminated |