Recall Z-1445-2013

I-Flow LLC · initiated June 1, 2012

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

I-Flow ON-Q Pain Relief System, T-Block Needle/Catheter Kit, Model #TB089T The intended use of each component in the finished final trays is consistent with the intended use of the component stated in the respective 510(k) submission.

Codes / lots: Lot # AW207202U

Reason for recall

I-Flow initiated two voluntary product removals of the ON-Q Pain Relief System: T-Block Needle/Catheter Tray and T-Block Needle/Catheter Kit because they may have potentially malfunctioned.

FDA-determined cause: Use error

Action taken

IFlow sent the customer notification letter on May 29, 2013, to 28 customers via FedEx along with a copy of the Service Guide to inform customers of the issue that can cause the Sting Ray to disconnect from the Catheter and how to prevent this problem from occuring. Customers are informed that in certain circumstances, ultrasound gel on the catheters significantly reduces the amount of force required to pull the catheter out of the Stingray connector. The In Service Guide clarifies that the method of catheter securement. It also instructs that wiping off any excess ultrasound gel from the catheter prior to inserting it into the Stingray connector will help with connecting the catheter to the Stingray connector. Customers are instructed that there is no action required for the I Flow ON Q Pain Relief System and ON Q T Bloc. The recall letter is only to provide further instructions on successful set up of the SPIROL catheter with the Stingray Connector included in the ON-Q T-bloc trays. Customers with any questions are instructed to call IFlow Customer service at (800) 448-3569 or IFlow 24 hr Tech support at (800) 444-2728. On June 1, 2012, an I-Flow sales representative removed the catheters and connectors from the one consignee/customers. For questions regarding this recall call 949-206-2700.

Details

Recall number
Z-1445-2013
Initiated
June 1, 2012
Posted by FDA
June 3, 2013
Terminated
June 25, 2013
Product code
BSO – Catheter, Conduction, Anesthetic
Quantity
75 units
Distribution
Nationwide distribution: USA including states of: CA, CT, FL, IL, IN, KY, NY, OH, OR, TN, TX, VA and WI.
Recalling firm
I-Flow LLC, Lake Forest, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1444-2013 Class III-Flow ON-Q Pain Relief System, T-Block Needle/Catheter Tray, Model #TBTB1089T The…I-Flow initiated two voluntary product removals of the ON-Q Pain Relief System: T-Block Needle/Catheter Tray… 06/01/2012Terminated