Recall Z-1445-2022

Aesculap Implant Systems LLC · initiated September 8, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

AESCULAP S4 Cervical Cross Connector, L26mm, S4C CROSS CONNECTOR FIXED 26MM, product code SW114T. Used in spinal fusion surgery.

Codes / lots: UDI-DI: 04038653362318

Reason for recall

Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.

FDA-determined cause: Device Design

Action taken

An Urgent Medical Device Removal Notification issued September 08, 2020 sent via USPS certified mail. The notice identifies the affected part numbers S4C CROSS CONNECTOR SW112T, SW113T, SW114T, SW115T, SW116T and SW117T and asks distributors and customers to identify existing inventory, quarantine, and return affected inventory. If further distributed, identify and notify customers. The notice includes an acknowledgement form to be completed and returned to Aesculap Inc. via email at qa-recalls@aesculapusa.com. Questions are to be directed to Lindsay Chromiak, Manager, QA Compliance and Vigilance (610) 984-9072.

Details

Recall number
Z-1445-2022
Initiated
September 8, 2020
Terminated
April 10, 2024
Product code
NKB – Thoracolumbosacral Pedicle Screw System
Quantity
24
Distribution
United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.
Recalling firm
Aesculap Implant Systems LLC, Center Valley, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1443-2022 Class IIAESCULAP S4 Cervical Cross Connector L22mm, S4C CROSS CONNECTOR FIXED 22MM, product code…Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement… 09/08/2020Terminated
Z-1444-2022 Class IIAESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code…Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement… 09/08/2020Terminated
Z-1446-2022 Class IIAESCULAP S4 Cervical Cross Connector L 28-33mm, S4C CROSS CONNECTOR VARIABLE 28-33MM…Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement… 09/08/2020Terminated
Z-1447-2022 Class IIAESCULAP S4 Cervical Cross Connector L 33-42mm, S4C CROSS CONNECTOR VARIABLE 33-42MM…Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement… 09/08/2020Terminated
Z-1448-2022 Class IIAESCULAP S4 Cervical Cross Connector L 42-58mm, S4C CROSS CONNECTOR VARIABLE 42-58MM…Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement… 09/08/2020Terminated