Recall Z-1451-2013

Maytex Corp · initiated May 15, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Comfort brand Ear-Loop Face Mask, Item #: 3100, Model Numbers: # 3100B, # 3100P, # 3100G, # 3100Y, # 3100W; Product is manufactured and distributed by Maytex Corporation, CA Ear-loop face mask for basic medical/dental procedures.

Codes / lots: All cases of item #: 3100 face mask with PO #. PO #, Date of receipt: 809-01, 2/5/08; 837-02, 7/29/08; 852-03, 10/10/08; 901-01, 2/3/09; 921-02, 6/8/09; 923-03, 7/8/09; 926-04, 8/26/09; 931-05, 11/10/09; 942-06, 12/16/09; 001-01, 2/17/10; 003-03, 4/12/10; 002-02 , 3/24/10; 003-03A, 7/23/10; 042-04, 11/29/10; 105-01, 1/31/11; 117-02, 4/29/11; 127/03, 7/8/11; 147-04, 12/16/11; 202-01, 3/8/2012; 224-02, 5/24/2012; 235-03, 9/12/2012; 253-04 , 12//192012

Reason for recall

Recent Lab testing conducted in March 2013 indicates that the product does not meet the standard of "Differential Pressure of only 1.6 Delta P (mm H2O/cm2)" as indicated on the box/package of the product; and also product manufactured between 2008 - 2012 cannot be guaranteed as Fluid Resistant.

FDA-determined cause: Labeling False and Misleading

Action taken

The firm, Maytex Corporation, sent a "MEDICAL DEVICE SAFETY NOTICE" dated May 15, 2013, to all its customers. The notice described the product, problem and actions to be taken. The customers were instructed to immediately contact your medical professional for treatment and contact Maytex Corporation to file a complaint should any illness or injury occur; discontinue use of the product if you detect any potential health hazards; if this product has been further sold or distributed, notify your customers to the retail level and send copies of safety notice to customer/users; and complete and promptly return the enclosed Medical Device Safety Notice Return Response Acknowledgement and Receipt form via mail to: Maytex Corporation 23521 Foley Street, Hayward CA 94545; fax at; 510-786-0209; email: www.maytex18@aol.com or by delivery to Maytex Corporation. Maytex Corporation has temporarily taken down their Website at www.maytexcorp.com to correct the wrong product specifications as described in the notice. If you have any questions, contact Maytex Corporation at 1-800-462-9839 or email: maytex18@aol.com Monday through Friday, 8:30 AM to 5:00 PM Pacific Time.

Details

Recall number
Z-1451-2013
Initiated
May 15, 2013
Posted by FDA
June 4, 2013
Terminated
October 29, 2013
Product code
FXX – Mask, Surgical
Quantity
43,915 Cases
Distribution
Worldwide distribution: USA (nationwide) including states of: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, MA, MD, ME, MI, MN, MO, MT, NE, NH, NJ, NM, NY, OK, OR, PA, PR, TN, TX, UT, WA, and WI; and country of: Mexico.
Recalling firm
Maytex Corp, Hayward, CA