Product
Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Simplify Disc Size 2, Ht 5/MD-5; Simplify Disc Size 2, Ht 5, 5/MD-5L; Simplify Disc Size 2, Ht 6/MD-6; Simplify Disc Size 2, Ht 6, 5/MD-6L; Simplify Disc Size 3, Ht 5/LG-5; Simplify Disc Size 3, Ht 5, 5/LG-5L; Simplify Disc Size 3, Ht 6/LG-6; Simplify Disc Size 3, Ht 6, 5/LG-6L
Codes / lots: Part/Correct GTIN (DI)/Lot: SM-5/843285100023/W79256, W82070, W80144; SM-6/843285100047/W80703, W81624, W80701, W81132, W81133, W81633, W81625, W81611, W81828, W81971, W82544, W82543, W81034, W81975, W81580, W81612, W81621, W81131, W81613, W80909, W81623, W81634, W79466, W81033, W81134, W81600, W79259; MD-5/843285100061/W81136, W81135, W80704, W80700; MD-5L/843285100078/W80707, W80706; MD-6/843285100085/W80911, W81642, W81643, W82545, W82546; MD-6L/843285100092/W81822; LG-5/843285100108/W81138, W81141, W79465, W81994, W81142, W81626, W81143, W79257; LG-5L/843285100115/W79258, W80146, W80915, W82548; LG-6/843285100122/W82547; LG-6L/843285100139/W81827
Reason for recall
Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.
FDA-determined cause: Labeling Change Control
Action taken
On 06/20/22, Field Safety notices were delivered in person, emailed, or mailed to hospital and distributor consignees who were informed of the following: 1) Complete and return the consignee confirmation form. 2) Return of affected devices is not required. However, if you would like these items exchanged, please contact your firm representative to request a Return Material Authorization number (RMA). 3) This recall notice should be shared with anyone who needs to be aware within your organization or shared with any organization where the potentially affected products have been transferred. Customers with questions, complaints, or who want to report adverse effects can contact the firm at complaints@nuvasive.com
Details
- Recall number
- Z-1453-2022
- Initiated
- June 20, 2022
- Product code
- MJO – Prosthesis, Intervertebral Disc
- Quantity
- 1,963
- Distribution
- US: CA, WA, TX, IN, MO, MI, FL, VA, OR. OUS: DE, NZ, SG, UK
- Recalling firm
- NuVasive Inc, San Diego, CA