Recall Z-1453-2022

NuVasive Inc · initiated June 20, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Simplify Disc Size 2, Ht 5/MD-5; Simplify Disc Size 2, Ht 5, 5/MD-5L; Simplify Disc Size 2, Ht 6/MD-6; Simplify Disc Size 2, Ht 6, 5/MD-6L; Simplify Disc Size 3, Ht 5/LG-5; Simplify Disc Size 3, Ht 5, 5/LG-5L; Simplify Disc Size 3, Ht 6/LG-6; Simplify Disc Size 3, Ht 6, 5/LG-6L

Codes / lots: Part/Correct GTIN (DI)/Lot: SM-5/843285100023/W79256, W82070, W80144; SM-6/843285100047/W80703, W81624, W80701, W81132, W81133, W81633, W81625, W81611, W81828, W81971, W82544, W82543, W81034, W81975, W81580, W81612, W81621, W81131, W81613, W80909, W81623, W81634, W79466, W81033, W81134, W81600, W79259; MD-5/843285100061/W81136, W81135, W80704, W80700; MD-5L/843285100078/W80707, W80706; MD-6/843285100085/W80911, W81642, W81643, W82545, W82546; MD-6L/843285100092/W81822; LG-5/843285100108/W81138, W81141, W79465, W81994, W81142, W81626, W81143, W79257; LG-5L/843285100115/W79258, W80146, W80915, W82548; LG-6/843285100122/W82547; LG-6L/843285100139/W81827

Reason for recall

Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.

FDA-determined cause: Labeling Change Control

Action taken

On 06/20/22, Field Safety notices were delivered in person, emailed, or mailed to hospital and distributor consignees who were informed of the following: 1) Complete and return the consignee confirmation form. 2) Return of affected devices is not required. However, if you would like these items exchanged, please contact your firm representative to request a Return Material Authorization number (RMA). 3) This recall notice should be shared with anyone who needs to be aware within your organization or shared with any organization where the potentially affected products have been transferred. Customers with questions, complaints, or who want to report adverse effects can contact the firm at complaints@nuvasive.com

Details

Recall number
Z-1453-2022
Initiated
June 20, 2022
Product code
MJO – Prosthesis, Intervertebral Disc
Quantity
1,963
Distribution
US: CA, WA, TX, IN, MO, MI, FL, VA, OR. OUS: DE, NZ, SG, UK
Recalling firm
NuVasive Inc, San Diego, CA