Recall Z-1454-2022

Abbott Laboratories · initiated June 9, 2022

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

REF: 4P52-20 Hemoglobin A1c for use with Architect, IVD, CE, R1 1 x 52 mL, R2 1 x 20 mL, A1cDL 2 x 35 mL

Codes / lots: UDI/DI: 00380740102746/ Lots: 59588UQ04, 59868UQ05, 59900UQ06, 60138UQ07, 60237UQ08, 60535UQ08, 60698UQ09, 60880UQ10, 61066UQ11, 61246UQ01, 61250UQ01

Reason for recall

There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.

FDA-determined cause: Under Investigation by firm

Action taken

On June 9, 2022, Abbott issued a "Urgent Medical Device Correction" to affected consignees via FedEx. 1. Immediately install the updated assay file versions listed below (as applicable for your laboratory), and manually configure the Low-Linearity value. The assay files can be obtained from www.corelaboratory.abbott 2. For the whole blood application, manually configure the THbWB Low-Linearity to 53.8278. For the hemolysate application, manually configure the THbH Low-Linearity to 1250. For detailed information on editing assay parameters, refer to Changing assay configuration settings, Change a linearity range in the ARCHITECT System Operations Manual, Section 2. 3. Please review this letter with your Medical Director or Laboratory Management and follow your laboratory protocol regarding the need for reviewing previously reported patient results. 4. Complete and return the Customer Reply Form. 5. Please retain this letter for your laboratory records. 6. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter.

Details

Recall number
Z-1454-2022
Initiated
June 9, 2022
Product code
PDJ – Hemoglobin A1c Test System
Quantity
103,316
Distribution
US: AK AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND NE NJ NV NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: ALBANIA ALGERIA ANDORRA ANGOLA ARGENTINA ARMENIA AUSTRALIA AUSTRIA AZERBAIJAN BAHAMAS BANGLADESH BELARUS BELGIUM BERMUDA BOSNIA AND HERZEGOVI BRAZIL BRUNEI BULGARIA BURKINA FASO CANADA CHILE CHINA COLOMBIA COSTA RICA Country CROATIA CYPRUS CZECH REPUBLIC…
Recalling firm
Abbott Laboratories, Irving, TX

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-1455-2022 Class IIREF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL…There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using… 06/09/2022Open, Classified
Z-1456-2022 Class IIREF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL…There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using… 06/09/2022Open, Classified