Recall Z-1461-2023

Baxter Healthcare Corporation · initiated March 27, 2023

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

GEM NEUROTUBE, Absorbable Woven Polyglycolic Acid Mesh Tube, Product Codes: a) 5311-01240-010, b) 5311-01240-012; for peripheral nerve repair

Codes / lots: a) 5311-01240-010, Lot Number 19012112, exp. 12/31/2023; b) 5311-01240-012, Lot Number 19012112, exp. 12/31/2023

Reason for recall

The product is brittle and potentially crumbling upon handling or when being removed from its package.

FDA-determined cause: Under Investigation by firm

Action taken

Baxter issued an Urgent Medical Device Recall notice to its consignees on 03/27/2023 by mail (USPS). The notice explained the issue and the risk and requested the following actions: 1. Immediately locate, isolate, and cease all use of the affected lot number of the product. 2. Contact Synovis MCA to arrange for return and replacement of the product. Synovis MCA can be reached at 800-510-3318 or 205-941-0111 between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday. Please have your Synovis MCA account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any inventory. Log in to the portal using the account number listed on the enclosed reply form instruction sheet. 4. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them.

Details

Recall number
Z-1461-2023
Initiated
March 27, 2023
Posted by FDA
April 25, 2023
Product code
JXI – Cuff, Nerve
Quantity
169 units
Distribution
Worldwide distribution - US Nationwide and the countries of China, Germany, Sweden, Norway, Netherlands, Greece, Switzerland, France
Recalling firm
Baxter Healthcare Corporation, Deerfield, IL